R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.5.5 CALCulate:LIMit:UPPer Subsystem<br />

The CALCulate:LIMit:UPPer subsystem defines the upper limit line.<br />

CALCulate Subsystem<br />

CALCulate:LIMit:UPPer[:DATA] ,...<br />

This command defines the values for the upper limit lines.<br />

Up to 8 limit lines can be defined at the same time. This command is independent of the measurement<br />

window.<br />

The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit line<br />

have to be identical. Otherwise default values are entered for missing values or not necessary values<br />

are deleted.<br />

The unit must be identical with the unit selected by CALC:LIM:UNIT. If no unit is indicated, the unit<br />

defined with CALC:LIM:UNIT is automatically used.<br />

In analyzer mode, the units DEG, RAD, S, HZ, PCT are not available.<br />

In analyzer mode, the limit check signals errors if the measured values exceed the UPPer limit line.<br />

Example: "CALC:LIM2:UPP -10,0,0,-10,-5"<br />

'Defines 5 upper limit values for limit line 2 in the preset unit.<br />

"CALC:LIM2:UPP?"<br />

'Outputs the upper limit values for limit line 2 separated by a comma.<br />

Characteristics: *RST value: - (LIMit:STATe is set to OFF)<br />

SCPI: conform<br />

Mode: A<br />

CALCulate:LIMit:UPPer:MARGin <br />

This command defines a margin to an upper limit line, at which out-of-limit values are signaled (if the<br />

limit check is active), but not handled as a violation of the limit value. The margin is independent of the<br />

measurement window.<br />

Example: "CALC:LIM2:UPP:MARG 10dB"<br />

'Defines the margin of limit line 2 to 10 dB below the limit value.<br />

Characteristics: *RST value: 0<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:UPPer:MODE RELative | ABSolute<br />

This command selects the relative or absolute scaling for the y-axis of the selected upper limit line. The<br />

setting is independent of the measurement window.<br />

Selecting RELative causes the unit to be switched to DB.<br />

Example: "CALC:LIM2:UPP:MODE REL"<br />

'Defines the y-axis of limit line 2 as relative scaled.<br />

Characteristics: *RST value: ABSolute<br />

SCPI: device-specific<br />

Mode: A<br />