R&S <strong>FSQ</strong> Instrument Functions<br />

SIGNAL<br />

COUNT<br />

REFERENCE<br />

FIXED<br />

<strong>Analyzer</strong> Mode<br />

• The marker measures only the position of the pixel on the trace and infers the<br />

frequency of the signal from this value. The trace, however, contains only a limited<br />

number of pixels. Depending upon the selected span, each pixel may contain<br />

many measurement values, which therefore limits the frequency resolution.<br />

• The resolution with which the frequency can be measured is proportional to the<br />

measurement time. For this reason, the bandwidth is normally made as wide as<br />

possible and the sweep time as short as possible. This results in a loss of<br />

frequency resolution.<br />

For the measurement with the frequency counter, the sweep is stopped at the reference<br />

marker, the frequency is counted with the desired resolution and then the<br />

sweep is allowed to continue.<br />

The SIGNAL COUNT softkey switches the frequency counter on/off.<br />

The frequency is counted at the position of the reference marker (MARKER 1). The<br />

sweep stops at the reference marker until the frequency counter has delivered a<br />

result. The time required for a frequency measurement depends on the selected frequency<br />

resolution. The resolution is set in the side menu.<br />

If no marker is enabled when the SIGNAL COUNT softkey is pressed, MARKER 1 is<br />

switched on and set at the largest signal.<br />

In addition, the SIGNAL COUNT function is displayed in the marker info field on the<br />

screen with [Tx CNT].<br />

Switching the SIGNAL COUNT function off is accomplished by pressing the softkey<br />

again.<br />

Remote command: CALC:MARK1:COUN ON;<br />

CALC:MARK:COUN:FREQ?<br />

The resolution of the frequency counter is set in the NEXT menu of the MARKER<br />

menu. R&S <strong>FSQ</strong> offers counter resolutions between 0.1 Hz and 10 kHz.<br />

REF FXD (ON OFF)<br />

REF POINT LEVEL<br />

REF POINT LVL OFFSET<br />

REF POINT FREQUENCY<br />

REF POINT TIME<br />

PEAK SEARCH<br />

The REFERENCE FIXED softkey defines the level and the frequency or time of<br />

MARKER 1 as a reference for one or several delta markers. The measured values<br />

for one or several markers displayed in the marker info field are derived from this<br />

reference point instead of the current values of the reference marker (MARKER 1).<br />

On actuating the softkey, reference fixed is switched on and thus, the level value<br />

and the frequency, time or x-level value of MARKER 1 immediately become the reference<br />

point.<br />

Operating Manual 1313.9681.12 - 01 4.57