R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe]:SWEep:IF:SHIFt:FREQuency <br />

SENSe Subsystem<br />

This command defines the signal frequency to automatically select the suitable 1st IF shifting for mode<br />

IF SHIFT AUTO (SENS:SWE:IF:SHIFT:MODE AUTO).<br />

Example: "SWE:IF:SHIF AUTO"<br />

’activates IF SHIFT AUTO.<br />

"SWE:IF:SHIF:FREQ 2GHz"<br />

’carrier signal at 2 GHz.<br />

Characteristics: *RST value: 1 GHz<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]SWEep:MODE AUTO | LIST | ESPectrum<br />

This command controls the sweep continuation.<br />

AUTO free running sweep - the spurious list is activated<br />

ESPectrum activates the Spectrum Emission Mask measurement<br />

LIST Activates the spurious emissions measurement. The sweep list is<br />

defined using the SENS:LIST:RANGE commands.<br />

Example: "SWE:MODE LIST"<br />

Activates the spurious emissions measurement.<br />

Characteristics: *RST value: AUTO<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]SWEep:POINts 155, 313, 625, 1251, 1999, 2501, 5001, 10001, 20001, 30001<br />

This command defines the number of measurement points for one sweep run.<br />

Example: "SWE:POIN 313"<br />

Characteristics: *RST value: 625<br />

SCPI: conform<br />

Mode: A<br />

[SENSe:]SWEep:TIME 2.5 ms to 16000 s (frequency domain) | 1 µs to 16000 s (time domain)<br />

This command defines the sweep time. The available time values are different in the frequency domain<br />

(2.5 ms to 16000 s with span > 0) and in the time domain (1 µs to 16000 s with span = 0).<br />

Used as a query, this command also returns the acquisition time of the FFT filter.<br />

If SWEep:TIME is directly programmed, automatic coupling to resolution bandwidth and video<br />

bandwidth is switched off.<br />

Example: "SWE:TIME 10s"<br />

Characteristics: *RST value: - (AUTO is set to ON)<br />

SCPI: conform<br />

Mode: A<br />