R&S <strong>FSQ</strong> Instrument Functions<br />

4.5.11.4 Measurement of the Filter or <strong>Signal</strong> Bandwidth<br />

<strong>Analyzer</strong> Mode<br />

N DB DOWN The N DB DOWN softkey activates the temporary markers T1 and T2 which are n<br />

dB below the active reference marker. Marker T1 is placed to the left and marker T2<br />

at the right of the reference marker. The value n can be input in a window.<br />

The default setting is 3 dB.<br />

Span > 0: The frequency spacing of the two temporary markers is indicated in the<br />

marker info field.<br />

Span = 0: The pulse width between the two temporary markers is indicated in the<br />

marker info field.<br />

PEAK LIST<br />

If, for example, it is not possible to form the frequency spacing for the n dB value<br />

because of the noise display, dashes are indicated instead of a measured value.<br />

If a negative value is entered than the markers are placed n dB above the active reference<br />

marker. This is then a n dB up function which can be used for notch filter<br />

measurements:<br />

To switch n dB down on or off:<br />

Remote command: CALC:MARK1:FUNC:NDBD:STAT ON<br />

To query the result pulse width:<br />

Remote command: CALC:MARK1:FUNC:NDBD:RES?<br />

To query the two marker x-values (in seconds) separated by comma:<br />

Remote command: CALC:MARK1:FUNC:NDBD:TIME? 'Span = 0<br />

Further remote commands:<br />

Remote command: CALC:MARK1:FUNC:NDBD 3dB<br />

CALC:MARK1:FUNC:NDBD:FREQ? 'Span > 0<br />

4.5.11.5 Measurement of a Peak List<br />

NEW SEARCH<br />

PEAK LIST (ON OFF)<br />

AUTOSEARCH (ON OFF)<br />

SORT MODE (FREQ LEVEL)<br />

PEAK EXCURSION<br />

SEARCH LIMITS LEFT LIMIT | RIGHT LIMIT<br />

MKR SYMBOL (ON OFF)<br />

NEXT PAGE<br />

MKR SYMBOL (ON OFF)<br />

MKR NUMBER (ON OFF)<br />

MAX PEAK COUNT (50)<br />

THRESHOLD<br />

SEARCH LIM OFF<br />