R&S <strong>FSQ</strong> Instrument Functions<br />

EDIT LIMIT<br />

LINE /<br />

NEW LIMIT<br />

LINE<br />

Basic Settings<br />

• the definition of the limit line as either upper or lower limit.<br />

• the data points for frequency/time and level.<br />

At the time of entry, the R&S <strong>FSQ</strong> immediately checks that all limit lines are in accordance<br />

with certain guidelines. These guidelines must be observed if specified operation<br />

is to be guaranteed.<br />

• The frequencies/times for each data point must be entered in ascending order,<br />

however, for any single frequency/time, two data points may be input (vertical<br />

segment of a limit line).<br />

The data points are allocated in order of ascending frequency/time. Gaps are not<br />

allowed. If gaps are desired, two separate limit lines must be defined and then<br />

both enabled.<br />

• The entered frequencies/times need not necessarily be selectable in R&S <strong>FSQ</strong>. A<br />

limit line may also exceed the specified frequency or time domains. The minimum<br />

frequency for a data point is -200 GHz, the maximum frequency is 200 GHz. For<br />

the time domain representation, negative times may also be entered. The valid<br />

range is -1000 s to +1000 s.<br />

• The minimum/maximum value for a limit line is -200 dB to +200 dB for the<br />

logarithmic or 10-20 to 10+20 or -99.9% to + 999.9% for the linear amplitude scales.<br />

NAME<br />

VALUES<br />

INSERT VALUE<br />

DELETE VALUE<br />

SHIFT X LIMIT LINE<br />

SHIFT Y LIMIT LINE<br />

SAVE LIMIT LINE<br />

The EDIT LIMIT LINE and NEW LIMIT LINE softkeys both call the EDIT LIMIT LINE<br />

submenu used for editing limit lines. In the table heading, the characteristics of the<br />

limit line can be entered. The data points for frequency/time and level values are<br />

entered in the columns.<br />

Name Enter name.<br />

Domain Select domain.<br />

Unit Select units.<br />

X-Axis Select interpolation<br />

Limit Select upper and lower limit value.<br />

X-Scaling Entry of absolute or relative values for the x-axis<br />

Y-Scaling Entry of absolute or relative values for the y-axis<br />

Margin Entry of margin.<br />

Threshold Entry of vertical threshold (only with relative vertical<br />

scaling)<br />

Comment Enter comments.<br />

Time/Frequency Enter time/frequency for the data points.<br />

Limit/dBm Enter magnitudes for the data points.<br />