R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate Subsystem<br />

"CALC:LIM:ACP:ACH:REL:STAT ON"<br />

'Switches on the check of the relative limit values for adjacent channels in<br />

screen A.<br />

"CALC:LIM:ACP:ACH:ABS:STAT ON"<br />

'Switches on the check of absolute limit values for the adjacent channels in<br />

screen A.<br />

"INIT;*WAI"<br />

'Starts a new measurement and waits for the sweep end.<br />

"CALC:LIM:ACP:ACH:RES?"<br />

'Queries the limit check result in the adjacent channels in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:ACPower:ACHannel:RESult?<br />

This command queries the result of the limit check for the upper/lower adjacent channel in the selected<br />

measurement window when adjacent channel power measurement is performed.<br />

If the power measurement of the adjacent channel is switched off, the command produces a query<br />

error.<br />

The numeric suffixes are irrelevant for this command.<br />

Parameter: The result is returned in the form , where<br />

= PASSED | FAILED, and where the first returned value denotes the<br />

lower, the second denotes the upper adjacent channel.<br />

Example: "CALC:LIM:ACP:ACH 30DB, 30DB"<br />

'Sets the relative limit value in screen A for the power in the lower and upper<br />

adjacent channel to 30 dB below the channel power.<br />

"CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM"<br />

'Sets the absolute limit value in screen A for the power in the lower and upper<br />

adjacent channel to -35 dB.<br />

"CALC:LIM:ACP ON"<br />

'Switches on globally the limit check for the channel/adjacent channel<br />

measurement in screen A.<br />

"CALC:LIM:ACP:ACH:STAT ON"<br />

'Switches on the limit check for the adjacent channels in screen A.<br />

"INIT;*WAI"<br />

'Starts a new measurement and waits for the sweep end.<br />

"CALC:LIM:ACP:ACH:RES?"<br />

'Queries the limit check result in the adjacent channels in screen A.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

This command is a query and therefore has no *RST value.<br />