R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Example: "CALC:MARK:FUNC:POW:SEL ACP"<br />

'Switches on adjacent channel power measurement in window A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-F<br />

CALCulate:MARKer:FUNCtion:POWer:STANdard:CATalog?<br />

CALCulate Subsystem<br />

This command returns all predefine ACP standards and all user standards. The numeric suffixes at<br />

MARKer are irrelevant for this command.<br />

Example: "CALC:MARK:FUNC:POW:STAN:CAT?”<br />

'get all available ACP standards<br />

Characteristics: *RST value:-<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:POWer:STANdard:DELete <br />

This command removes the ACP user standard . It is not possible to remove predefine ACP<br />

standards. The numeric suffixes at MARKer are irrelevant for this command.<br />

Example: "CALC:MARK:FUNC:POW:SEL ACP"<br />

' activate ACP measurement<br />

"CALC:MARK:FUNC:POW:PRES FW3G"<br />

' use WCDMA 3GPP as basis<br />

"CALC:LIM:ACP:ACH:REL -60,-60"<br />

' change ACP limit to -60dBc<br />

"CALC:MARK:FUNC:POW:STAN:SAVE 'my_acp_std"<br />

' save as ACP USER standard my_acp_std<br />

"CALC:MARK:FUNC:POW:STAN:DEL 'my_acp_std"<br />

' delete the user standard my_acp_std<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:POWer:STANdard:SAVE <br />

This command saves the current ACP settings as an ACP user standard. The numeric suffixes at<br />

MARKer are irrelevant for this command.<br />

Example: "CALC:MARK:FUNC:POW:SEL ACP"<br />

' activate ACP measurement<br />

"CALC:MARK:FUNC:POW:PRES FW3G"<br />

' use WCDMA 3GPP as basis<br />

"CALC:LIM:ACP:ACH:REL -60,-60"<br />

' change ACP limit to -60dBc<br />

"CALC:MARK:FUNC:POW:STAN:SAVE 'my_acp_std"<br />

' save as ACP user standard my_acp_std<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-F<br />

Operating Manual 1313.9681.12 - 01 6.77