R&S <strong>FSQ</strong> Instrument Functions<br />

<strong>Analyzer</strong> Mode<br />

X-AXIS RANGE The X-AXIS RANGE softkey changes the level range to be covered by the statistics<br />

measurement selected.<br />

The function is identical to softkey RANGE LOG MANUAL in AMPT menu.<br />

Remote command: CALC:STAT:SCAL:X:RANG <br />

Y-UNIT (% ABS) The softkey Y-UNIT (% ABS) defines the scaling type on the y-axis. The default<br />

case are the absolute probability. This can be changed to percent values. The softkeys<br />

Y-AXIS MIN and Y-AXIS MAX are using values based on the Y-UNIT setting.<br />

Y-AXIS MAX<br />

VALUE<br />

Y-AXIS MIN<br />

VALUE<br />

ADJUST<br />

SETTINGS<br />

DEFAULT<br />

SETTINGS<br />

Remote command: CALC:STAT:SCAL:Y:UNIT PCT | ABS<br />

The 0.01%, 0.1%, 1% and 10% value from the CCDF measurement are shown in<br />

the bottom screen half. Those values can also queried via remote:<br />

Remote command: CALC:STAT:CCDF:X? P0_01 | P0_1 |<br />

P1 | P10<br />

The Y-AXIS MAX VALUE softkey defines the upper limit of the displayed probability<br />

range.<br />

Values on the y-axis are normalized which means that the maximum value is 1.0. As<br />

the y-axis scaling has a logarithmic axis the distance between max and min value<br />

must be at least one decade.<br />

Remote command: CALC:STAT:SCAL:Y:UPP <br />

The Y-AXIS MIN VALUE softkey defines the lower limit of the displayed probability<br />

range.<br />

As the y-axis scaling has a logarithmic axis the distance between max and min value<br />

must be at least one decade. Valid values are in the range 0 < value < 1.<br />

Remote command: CALC:STAT:SCAL:Y:LOW <br />

The ADJUST SETTINGS softkey optimizes the level settings of the R&S <strong>FSQ</strong><br />

according to the measured peak power in order to gain maximum sensitivity of the<br />

instrument.<br />

The level range is adjusted according to the measured difference between peak and<br />

minimum power for APD measurement and peak and mean power for CCDF measurement<br />

in order to obtain maximum power resolution.<br />

Remote command: CALC:STAT:SCAL:AUTO ONCE<br />

The DEFAULT SETTINGS softkey resets the x- and y-axis scalings to their PRESET<br />

values.<br />

• x-axis ref level: -20 dBm<br />

• x-axis range APD: 100 dB<br />

• x-axis range CCDF: 20 dB<br />

• y-axis upper limit: 1.0<br />

• y-axis lower limit: 1E-6<br />

Remote command: CALC:STAT:PRES<br />

Operating Manual 1313.9681.12 - 01 4.113