R&S <strong>FSQ</strong> Instrument Functions<br />

SELECT<br />

MARKER<br />

TOI MKR CALC<br />

SRCH<br />

HARMONIC<br />

DISTOR<br />

<strong>Analyzer</strong> Mode<br />

The R&S <strong>FSQ</strong> calculates the third order intercept from the level spacing between<br />

normal markers and delta markers and outputs it in the marker info field.<br />

Remote command: CALC:MARK:FUNC:TOI ON;<br />

CALC:MARK:FUNC:TOI:RES?<br />

Example<br />

A two-tone signal with frequencies of 100 MHz and 101 MHz is applied to the RF<br />

input of the R&S <strong>FSQ</strong>. The level of the two signals is -10 dBm.<br />

[PRESET] The R&S <strong>FSQ</strong> is set to the default setting.<br />

[CENTER: 100.5 MHz] The center frequency is set to 100.5 MHz.<br />

[SPAN: 3 MHz] The span is set to 3 MHz.<br />

[AMPT: -10 dBm] The reference level is set to -10 dBm.<br />

[MKR FCTN] MARKER 1 is switched on and set to the signal peak.<br />

[TOI] The R&S <strong>FSQ</strong> sets the 4 markers to the useful signals and<br />

the intermodulation products and calculates the third order<br />

intercept. The result is output in the marker info field.<br />

The SELECT MARKER softkey activates the selection of a marker for functions<br />

MODULATION DEPTH and TOI. Thus, the markers can be fine-adjusted for these<br />

functions.<br />

The markers are numerically selected in a data entry field. Delta marker 1 is<br />

selected by entering '0'.<br />

If the marker is in the switch-off state, it will be switched on and can thus be shifted.<br />

Remote command: CALC:MARK1 ON;<br />

CALC:MARK1:X ;<br />

CALC:MARK1:Y?<br />

The TOI measurement is based on positioning the markers 3 and 4 onto the position<br />

of the calculated intermodulation products. The accuracy of these frequency calculations<br />

depend on the precision of the two carrier frequencies found with peak search.<br />

The softkey TOI MKR CALC SRCH allows to switch between the calculated marker<br />

position (CALC) and perfoming a local peak search near the expected frequencies<br />

(SRCH) instead of the faster position calculation.<br />

By default, the faster CALC method is used.<br />

Remote command: CALC:MARK:FUNC:TOI:MARK CALC<br />

4.5.13.9 Harmonic Measurement<br />

HARMONIC (ON OFF)<br />

NO. OF HARMONICS<br />

HARMONIC SWEEPTIME<br />

HARMONIC RBW AUTO<br />

ADJUST SETTINGS<br />

4.124 Operating Manual 1313.9681.12 - 01