R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

INPut:COUPling AC | DC<br />

This command switches the input coupling of the RF input between AC and DC.<br />

Example: "INP:COUP DC"<br />

Characteristics: *RST value: AC<br />

SCPI: conform<br />

Mode: A<br />

INPut:EATT 0 to 30dB<br />

INPut Subsystem<br />

This command programs the attenuation of the electronic input attenuator. The attenuation can be<br />

varied in 5 dB steps from 0 to 30 dB. Other entries are rounded to the next lower integer value. If the<br />

attenuation is programmed directly, the coupling to the reference level is switched off.<br />

If the defined reference level cannot be set with the given RF attenuation, this level is adapted to the<br />

maximum possible value.<br />

The electronic attenuator is switched off in the default state.<br />

Example: "INP:EATT:STAT ON"<br />

'Switches the electronic attenuator into the signal path.<br />

"INP:EATT 15dB"<br />

'Sets the attenuation of the electronic attenuator to 15 dB and switches off the<br />

coupling to the reference level.<br />

Characteristics: *RST value: 0 dB (state is set to OFF)<br />

SCPI: device-specific<br />

Mode: A<br />

The command is only available with the electronic attenuator option B25.<br />

INPut:EATT:AUTO ON | OFF<br />

This command automatically couples the electronic input attenuation to the reference level and the<br />

attenuation of the mechanical attenuator (state ON) or switches the input attenuation to manual entry<br />

(state OFF).<br />

Example: "INP:EATT:STAT ON"<br />

'Switches the electronic attenuator into the signal path.<br />

"INP:EATT:AUTO ON"<br />

'Couples the attenuation of the electronic attenuator to the reference level.<br />

Characteristics: *RST value: ON<br />

SCPI: device-specific<br />

Mode: A<br />

The softkey is only available for models 3, 8 and 26.<br />

The command is only available with the electronic attenuator option B25.<br />

Operating Manual 1313.9681.12 - 01 6.125