R&S <strong>FSQ</strong> Instrument Functions<br />

ACP REF<br />

SETTINGS<br />

• Even number of TX channels:<br />

<strong>Analyzer</strong> Mode<br />

The two TX channels in the middle are used to calculate the frequency between<br />

those two channels. This frequency is aligned to the center frequency.<br />

Adjacent channels<br />

Since all the adjacent channels often have the same distance to each other, the<br />

entry of the adjacent-channel spacing (ADJ) causes channel spacing ALT1 to be set<br />

to twice and channel spacing ALT2 to three times the adjacent-channel spacing (and<br />

so on). Thus only one value needs to be entered in case of equal channel spacing.<br />

The same holds true for the ALT2 channels when the bandwidth of the ALT1 channel<br />

is entered.<br />

The channel spacings can be set separately by overwriting the table from top to<br />

bottom.<br />

If the ACP or MCACP measurement is started, all settings according to the standard<br />

including the channel bandwidths and channel spacings are set and can be<br />

adjusted afterwards.<br />

Remote command: SENS:POW:ACH:SPAC:CHAN 20kHz<br />

SENS:POW:ACH:SPAC:ACH 20kHz<br />

SENS:POW:ACH:SPAC:ALT1 40kHz<br />

SENS:POW:ACH:SPAC:ALT2 60kHz<br />

...<br />

The ACP REF SETTINGS softkey opens a table for selecting the transmission channel<br />

to which the adjacent-channel relative power values should be referenced.<br />

4.98 Operating Manual 1313.9681.12 - 01