18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.12.1 INPut:DIQ Subsystem<br />

The INPut:DIQ subsystem controls the digital baseband input.<br />

This subsystem is available only with option R&S <strong>FSQ</strong>-B17.<br />

INPut Subsystem<br />

INPut:DIQ:CDEVice?<br />

This command queries the current configuration and status of the Digital Baseband input of option<br />

R&S <strong>FSQ</strong>-B17. The returned string includes the name of the connected device, the serial number of<br />

this device, the port name and the sample rate, separated by comma, e.g. “AMU200A,100266,Out A,<br />

80000000.0”.<br />

The suffix is irrelevant.<br />

Example: "INP:DIQ:CDEV?"<br />

queries the current input configuration.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is a query and therefore has no *RST value.<br />

INPut:DIQ:RANGe:AUTO ON | OFF<br />

This command controls the auto set function of the digital baseband input full scale value.<br />

Depending on the digital baseband signal source, the full scale value of the I/Q data is passed to the<br />

R&S <strong>FSQ</strong> LVDS input interface.<br />

If the full scale level of the signal source is outside the R&S <strong>FSQ</strong> supported range, it shows the “BDI”<br />

enhancement label. The “BDI” label is also visible if the signal source does not support automatic<br />

matching.<br />

The suffix is irrelevant.<br />

The command is available with option R&S <strong>FSQ</strong>-B17.<br />

Parameter: ON<br />

Automatically matches the full scale value of the digital baseband input to that of<br />

the signal source.<br />

OFF<br />

The R&S <strong>FSQ</strong> does not match the full scale value. You can set it manually with<br />

“INPut:DIQ:RANGe[:UPPer]”.<br />

Example: "INP:DIQ:RANG:AUTO OFF"<br />

disbales automatic matching of the full scale value.<br />

"INP:DIQ:RANG 1.2"<br />

sets the voltage to 1.2 V.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

6.128 Operating Manual 1313.9681.12 - 01