R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]CORRection:RECall<br />

SENSe Subsystem<br />

This command restores the instrument setting that was valid for the measurement of the reference<br />

data, provided that the tracking generator is active.<br />

Example: "CORR:REC"<br />

Characteristics: *RST value: -<br />

SCPI: conform<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

This command is only valid in conjunction with the ext. generator control option (B10).<br />

[SENSe:]CORRection[:STATe] ON | OFF<br />

This command activates/deactivates the normalization of the measurement results in the selected<br />

window provided that the tracking generator is active. The command is available only after acquisition<br />

of a reference trace for the selected type of measurement (transmission/reflection, see command<br />

[SENSe:]CORRection:COLLect[:ACQire]).<br />

Example: "CORR ON "<br />

'Activates normalization in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: conform<br />

Mode: A<br />

This command is only valid in conjunction with the ext. generator control option (B10).<br />

[SENSe:]CORRection:TRANsducer:ACTive?<br />

This command queries the active transducer factor. If no transducer factor is active, an empty string is<br />

returned.<br />

Example: "CORR:TRAN:ACT?"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is a query and therefore has no *RST value.<br />

[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe] ON | OFF<br />

This command switches the automatic adjustment of the reference level to the selected transducer<br />

factor on or off. Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.<br />

Example: "CORR:TRAN:SEL 'FACTOR1'"<br />

"CORR"CORR:TRAN:ADJ:RLEV ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.157