R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.6.6 CALCulate:MARKer:FUNCtion:STRack Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:MARKer:FUNCtion:STRack subsystem defines the settings of the signal track.<br />

CALCulate:MARKer:FUNCtion:STRack:BANDwidth|BWIDth 10Hz to MAX(SPAN)<br />

These commands have the same function. For the selected measurement window they define the<br />

bandwidth around the center frequency within which the largest signal is searched. The function is<br />

independent of the selected marker, i.e. the numeric suffix of MARKer is irrelevant. It is only<br />

available in the frequency domain (span > 0).<br />

Example: "CALC:MARK:FUNC:STR:BAND 1MHZ"<br />

'Sets the search bandwidth for screen A to 1 MHz.<br />

"CALC:MARK:FUNC:STR:BWID 1MHZ"<br />

'Alternative command for the same function.<br />

Characteristics: *RST value: -- (= span/10 on activating the function)<br />

SCPI: device-specific<br />

Mode: A-F<br />

CALCulate:MARKer:FUNCtion:STRack[:STATe] ON | OFF<br />

This command switches the signal-track function on or off for the selected measurement window. The<br />

function is independent of the selected marker, i.e. the numeric suffix of MARKer is irrelevant.<br />

With signal track activated, the maximum signal is determined after each frequency sweep and the<br />

center frequency is set to the frequency of this signal. Thus with drifting signals the center frequency<br />

follows the signal.<br />

Example: "CALC:MARK:FUNC:STR ON"<br />

'Switches on the signal track function for screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A-F<br />

CALCulate:MARKer:FUNCtion:STRack:THReshold -330dBm to +30dBm<br />

This command defines the threshold above which the largest signal is searched for in the selected<br />

measurement window. The function is independent of the selected marker, i.e. the numeric suffix<br />

of MARKer is irrelevant. It is only available in the frequency domain (span > 0).<br />

Note that the definition of the search bandwidth is possible only if the signal track function is on (CALC:<br />

MARK:FUNC:STR ON).<br />

The response unit depends on the settings defined with CALC:UNIT.<br />

Example: "CALC:MARK:FUNC:STR:THR -50DBM"<br />

'Sets the threshold for signal tracking in screen A to -50 dBm.<br />

Characteristics: *RST value: -120 dBm<br />

SCPI: device-specific<br />

Mode: A-F<br />

The entry of the search bandwidth is only possible if the <strong>Signal</strong> Track function is<br />

switched on (CALC:MARK:FUNC:STR ON).<br />

Operating Manual 1313.9681.12 - 01 6.79