R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:MARKer:FUNCtion:DEModulation[:STATe] ON | OFF<br />

CALCulate Subsystem<br />

This command switches on or off the audio demodulator when the indicated marker is reached in the<br />

selected measurement window. In the frequency domain (span > 0) the hold time can be defined at<br />

the corresponding marker position with CALCulate:MARKer:FUNCtion: DEModulation:HOLD.<br />

In the time domain (span = 0) the demodulation is permanently active.<br />

Example: "CALC2:MARK3:FUNC:DEM ON"<br />

'Switches on the demodulation for marker 3 in screen B.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:MDEPth:RESult?<br />

This command queries the AM modulation depth in the indicated measurement window.<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

The numeric suffix of :MARKer is irrelevant for this command.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK:X 10MHZ"<br />

'Sets the reference marker (marker 1) to the carrier signal at 10 MHz.<br />

"CALC:MARK:FUNC:MDEP ON"<br />

'Switches on the modulation depth measurement in screen A.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

"CALC:MARK:FUNC:MDEP:RES?"<br />

'Outputs the measured value of screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only a query and therefore has no *RST value.<br />

CALCulate:MARKer:FUNCtion:MDEPth[:STATe]<br />

This command switches on the measurement of the AM modulation depth. An AM-modulated carrier<br />

is required on the screen for correct operation. If necessary, marker 1 is previously activated and set<br />

to the largest signal available.<br />

The level value of marker 1 is regarded as the carrier level. On activating the function, marker 2 and<br />

marker 3 are automatically set as delta markers symmetrically to the carrier to the adjacent maxima of<br />

the trace.<br />

If the position of delta marker 2 is changed, delta marker 3 is moved symmetrically with respect to the<br />

reference marker (marker 1). If the position of delta marker 3 is changed, fine adjustment can be<br />

performed independently of delta marker 2.<br />

The R&S <strong>FSQ</strong> calculates the power at the marker positions from the measured levels.<br />

6.56 Operating Manual 1313.9681.12 - 01