R&S <strong>FSQ</strong> Instrument Functions<br />

EXPORT /<br />

IMPORT DEV<br />

DATA<br />

Basic Settings<br />

The EXPORT DEV DATA softkey opens a dialog to export device specific data to a<br />

memory stick.<br />

The export function automatically creates a sub folder using the instrument name<br />

and the serial number, e.g. <strong>FSQ</strong>40_123000 and copies device specific files to this<br />

folder:<br />

• Option license key codes<br />

• Board specific data<br />

• Frequency response correction file (option <strong>FSQ</strong>-B72 only)<br />

IMPORT DEV DATA reads these data back onto the instrument.<br />

Select the folder, where the instrument specific sub folder is located and press<br />

OPEN. A dialog asks to perform a reboot when the import is done.<br />

This function is helpful to backup device specific data, e.g. the option license keys. It<br />

is also helpful if you have to use a removable flash disk (R&S <strong>FSQ</strong>-B18) for more<br />

than one device. In this case:<br />

• Export the instrument data of device 1.<br />

• Switch instrument 1 off.<br />

• Remove the flash disk from device 1.<br />

• Insert the flash disk into device 2.<br />

• Switch instrument 2 on<br />

• Import the instrument data of device 2 (saved before).<br />

Note that a new export of the specific instrument data is required whenever:<br />

• A new option key was inserted.<br />

• The <strong>Rohde</strong> & <strong>Schwarz</strong> Service refitted a board.<br />

• The <strong>Rohde</strong> & <strong>Schwarz</strong> Service performed a device calibration.<br />

Remote command: ---<br />

Selftest<br />

SELFTEST The SELFTEST softkey initiates the selftest of the instrument modules.<br />

With this function the instrument is capable of identifying a defective module in case<br />

of failure.<br />

During the selftest a message box appears in which the current test and its result is<br />

shown. The test sequence can be aborted by pressing ENTER ABORT.<br />

All modules are checked consecutively and the test result (selftest PASSED or<br />

FAILED) is output in the message box.<br />

SELFTEST<br />

RESULTS<br />

Remote command: *TST?<br />

The SELFTEST RESULTS softkey calls the SELFTEST table in which the results of<br />

the module test are displayed.<br />