18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.6.4 CALCulate:MARKer:FUNCtion:HARMonics Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:MARKer:FUNCtion:HARMonics subsystem contains the commands for Harmonic Distortion<br />

measurement<br />

CALCulate:MARKer:FUNCtion:HARMonics:BANDwidth:AUTO ON | OFF<br />

This command specifies whether the resolution bandwidth of the 2nd to the nth harmonic should be<br />

identical to the bandwidth of the first harmonic (OFF) or to the next largest bandwidth (corresponding<br />

to the harmonic) of the multiple of the bandwidth of the first harmonic (ON).<br />

Example: "CALC:MARK:FUNC:HARM:BAND:AUTO OFF"<br />

'Switches automatic bandwidth expansion off.<br />

Characteristics: *RST value: ON<br />

SCPI: device-specific<br />

Mode: A-F, A-T<br />

CALCulate:MARKer:FUNCtion:HARMonics:BANDwidth[:LIST]?<br />

This command returns the resolution bandwidth values used by the harmonic measurment. A separate<br />

value for every harmonic frequency is automatically calculated. It is possible to query the calculated<br />

values with this command.<br />

The function is independent of the marker selection, i.e. the suffix or of CALCulate or<br />

MARKer is irrelevant.<br />

Example: "CALC:MARK:FUNC:HARM:BAND?"<br />

' Returns the used bandwidth values of the harmonics measurement.<br />

Characteristics: *RST value:-<br />

SCPI: device-specific<br />

This command is a query and therefore has no *RST value.<br />

CALCulate:MARKer:FUNCtion:HARMonics:DISTortion? TOTal<br />

This command queries the results of the total harmonic distortion (THD).<br />

A complete sweep with synchronization to sweep end should be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in the<br />

single-sweep mode.<br />

The function is independent of the marker selection, i.e. the suffix or of CALCulate<br />

or MARKer is irrelevant.<br />

Example: "INIT:CONT OFF"<br />

’Switches to single-sweep mode.<br />

"CALC:MARK:FUNC:HARM:NHARM 3"<br />

'Sets the number of harmonics to 3.<br />

"CALC:MARK:FUNC:HARM ON"<br />

’Switches on the measurement of harmonics.<br />

"INIT;*WAI"<br />

’Starts a sweep and waits for the end.<br />

"CALC:MARK:FUNC:HARM:DIST? TOT"<br />

’Outputs the total harmonic distortion in % and dB.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A-F, A-T<br />

6.68 Operating Manual 1313.9681.12 - 01

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!