R&S <strong>FSQ</strong> Instrument Functions<br />

ADJ REFLVL<br />

OFFSET<br />

<strong>Analyzer</strong> Mode<br />

Defines an additional level offset to the measured mean power to be taken into<br />

account for the analyzer's reference level setting.<br />

It is a modification to the ADJUST REF LEVEL function specific to user standards.<br />

Proceed as follows to adjust the offset value for a predefined standard:<br />

➢ Select a predefine standard with the CP/ACP STANDARD<br />

➢ Create a new user standards with SAVE AS USER STD<br />

➢ Select the user standard you just created with SELECT USER STD<br />

➢ Change the offset<br />

➢ Save the user standard again to keep the current settings.<br />

Remote command: POW:ACH:PRES:RLEV:OFFS 10dB<br />

SELECT TRACE The SELECT TRACE softkey selects the trace on which the CP/ACP measurement<br />

is to be performed. Only activated traces can be selected, i.e. traces not set to<br />

BLANK.<br />

Remote command: SENS:POW:TRAC 1<br />

Measurement of adjacent-channel power for a specific standard:<br />

The adjacent-channel power is to be measured for a signal at 800 MHz with 0 dBm<br />

level in line with IS136.<br />

[PRESET] Set the R&S <strong>FSQ</strong> to the default setting.<br />

[FREQ: CENTER: Set the center frequency to 800 MHz.<br />

800 MHz]<br />

[AMPT: 0 dBm] Set the reference level to 0 dBm.<br />

[MEAS] Call the menu for the measurement functions.<br />

[CHAN PWR / ACP] Select the channel and adjacent-channel power measurement<br />

function. The measurement is performed with the<br />

default settings or a previously defined setting. The submenu<br />

for setting the desired new configuration is opened.<br />

[CP/ACP STANDARD: Select the NADC (IS136) standard.<br />

select IS136: ENTER]<br />

[CP/ACP CONFIG] Call the submenu for configuration of the adjacent-channel<br />

power measurement.<br />

[NO. OF ADJ CHAN: 2 Select two adjacent channels for the measurement, i.e.<br />

ENTER]<br />

the adjacent channel and the alternate channel are measured.<br />

[ADJUST SETTINGS] Set the optimum span, resolution bandwidth (RBW), video<br />

bandwidth (VBW) and detector automatically for the measurement.<br />

The absolute channel power and the relative<br />

power of the adjacent channels are displayed on the<br />

screen.<br />

Change to the main menu for channel power measure-<br />

PREV<br />

ment.<br />

[ADJUST REF LVL] Set the reference level equal to the channel power measured.<br />

Operating Manual 1313.9681.12 - 01 4.103