R&S <strong>FSQ</strong> Instrument Functions<br />

EDIT TRD<br />

FACTOR<br />

Basic Settings<br />

The EDIT TRD FACTOR softkey gives access to the submenu for editing and generating<br />

transducer factors. For details refer to section “Entry and Editing of Transducer<br />

Factors” on page 4.179.<br />

NEW FACTOR The NEW FACTOR softkey gives access to the submenu for editing and generating<br />

transducer factors. For details refer to section “Entry and Editing of Transducer Factors”<br />

on page 4.179.<br />

DELETE<br />

FACTOR<br />

VIEW<br />

TRANSDUCER<br />

REFLVL ADJ<br />

AUTO MAN<br />

The DELETE FACTOR softkey deletes the marked factor.<br />

To prevent deletion by mistake, deletion has to be confirmed.<br />

Remote command: CORR:TRAN DEL<br />

The VIEW TRANSDUCER softkey activates the display of the active transducer factor<br />

or set. During the measurement, the interpolated characteristic of the active<br />

transducer factor or set is shown in the diagram instead of the measured values.<br />

Remote command: CORR:TRAN:VIEW ON<br />

When a transducer factor is used, the trace is moved by a calculated shift. However,<br />

an upward shift reduces the dynamic range for the displayed values. With the softkey<br />

REFLVL ADJ an automatic reference level offset adaptation allows to restore<br />

the original dynamic range by also shifting the reference level by the maximum<br />

value of the transducer factor.<br />

If transducers are active the function is automatically set to AUTO to obtain the best<br />

dynamic performance.<br />

Remote command: CORR:TRAN:SEL 'FACTOR1'<br />

CORR:TRAN:ADJ:RLEV ON | OFF<br />

PAGE UP The PAGE UP softkey is used to scroll through large tables that cannot completely<br />

be displayed on the screen.<br />

PAGE DOWN The PAGE DOWN softkey is used to scroll through large tables that cannot completely<br />

be displayed on the screen.<br />

Entry and Editing of Transducer Factors<br />

A transducer factor is characterized by the following:<br />

• Reference values with frequency and factor (Values)<br />

• Unit of the factor (Unit) and<br />

• Name (Name) to distinguish the various factors.<br />

During entry the R&S <strong>FSQ</strong> checks the transducer factor for compliance with specific<br />

rules that must be met to ensure correct operation.<br />

• The frequencies for the reference values must always be entered in ascending<br />

order. Otherwise the entry will not be accepted and the following message is<br />

displayed:<br />

WRONG FREQUENCY SEQUENCE !<br />

Operating Manual 1313.9681.12 - 01 4.179