R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.6 CALCulate:MARKer Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:MARKer subsystem checks the marker functions in the instrument.<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

The subsystem is divided into the description of the general marker commands (“General CALCulate:<br />

MARKer:... Commands” on page 6.45) and of the marker functions (“CALCulate:MARKer:FUNCtion<br />

Subsystem” on page 6.54). The description of marker functions is subdivided again into the commands<br />

for transient measurement (“CALCulate:MARKer:FUNCtion:ADEMod Subsystem” on page 6.95), the<br />

commands for harmonic distortion measurement (“CALCulate:MARKer:FUNCtion:HARMonics Subsystem”<br />

on page 6.68), for control of power measurement(“CALCulate:MARKer:FUNCtion:POWer Subsystem”<br />

on page 6.71), for settings of signal track (“CALCulate:MARKer:FUNCtion:STRack Subsystem”<br />

on page 6.79) and for controlling the main power functions (“CALCulate:MARKer:FUNCtion:SUMMary<br />

Subsystem” on page 6.81).<br />

6.5.6.1 General CALCulate:MARKer:... Commands<br />

CALCulate:MARKer:AOFF<br />

This command switches off all active markers in the indicated measurement window.<br />

All delta markers and active marker/delta marker measurement functions are also switched off.<br />

Example: "CALC:MARK:AOFF"<br />

'Switches off all markers in screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:MARKer:COUNt ON | OFF<br />

This command switches on or off the frequency counter at the marker position in the selected<br />

measurement window. The count result is queried with CALCulate:MARKer:COUNt:FREQuency?.<br />

Frequency counting is possible only for marker 1 in every measurement window. If it is activated for<br />

another marker, it is automatically de-activated for the previous marker.<br />

It should be noted that a complete sweep must be performed after switching on the frequency counter<br />

to ensure that the frequency to be measured is actually reached. The synchronization to the sweep<br />

end required for this is possible only in single-sweep mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK ON"<br />

'Switches on marker 1 in screen A.<br />

"CALC:MARK:COUN ON"<br />

'Switches on the frequency counter for marker 1.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

"CALC:MARK:COUN:FREQ?"<br />

'Outputs the measured value in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.45