R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.1 CALCulate:DELTamarker Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:DELTamarker subsystem controls the delta-marker functions in the instrument.<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

CALCulate:DELTamarker:AOFF<br />

This command switches off all active delta markers.<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

Example: "CALC:DELT:AOFF"<br />

'Switches off all delta markers in screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]<br />

<br />

This command sets the reference point level for all delta markers in the selected measurement window<br />

for a measurement with fixed reference point (CALC:DELT:FUNC:FIX:STAT ON) to the peak of the<br />

selected trace.<br />

For phase noise measurements (CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON), the<br />

command defines a new reference point level for delta marker 2 in the selected measurement window.<br />

Example: "CALC:DELT:FUNC:FIX:RPO:MAX"<br />

'Sets the reference point level for the delta marker in screen A to the peak of the<br />

trace.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:X <br />

This command defines a new reference frequency (span > 0) or time (span = 0) for all delta markers<br />

in the selected measurement window for a measurement with fixed reference value (CALCulate:<br />

DELTamarker:FUNCtion:FIXed:STATe ON).<br />

For phase noise measurements (CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON), the<br />

command defines a new reference frequency or time for delta marker 2 in the selected measurement<br />

window.<br />

Example: "CALC2:DELT:FUNC:FIX:RPO:X 128MHz"<br />

'Sets the reference frequency in screen B to 128 MHz.<br />

Characteristics: *RST value: - (FUNction:FIXed[:STATe] is set to OFF)<br />

SCPI: device-specific<br />

Mode: A<br />