R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

DIAGnostic:SERVice:INPut:PULSed[:STATe] ON | OFF<br />

DIAGnostic Subsystem<br />

This command toggles the calibration signal in the selected measurement window between pulsed and<br />

non-pulsed.<br />

The selection takes effect only if the RF input has been set to the internal reference signal using the<br />

DIAGnostic:SERVice:INPut[:SELect] command.<br />

Example: "DIAG:SERV:INP CAL"<br />

"DIAG:SERV:INP:PULS ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

DIAGnostic:SERVice:INPut:RECTangle:PRATe <br />

This command sets the frequency of the rectangle calibration signal in the selected measurement<br />

window (suffix ).<br />

5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz can be set.<br />

Example: "DIAG:SERV:INP:RECT:PRAT 128 MHz"<br />

Characteristics: *RST value: 5 kHz<br />

SCPI: device-specific<br />

Mode: A<br />

DIAGnostic:SERVice:INPut:RECTangle[:STATe] ON | OFF<br />

This command switches the rectangle calibration signal on or off in the selected measurement window<br />

(suffix ). The setting takes effect only if the RF input is set to the internal reference signal (DIAG:<br />

SERV:INP CAL command).<br />

Example: "DIAG:SERV:INP CAL"<br />

"DIAG:SERV:INP:RECT ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

DIAGnostic:SERVice:INPut[:SELect] CALibration | RF<br />

This command toggles between the RF input on the front panel and the internal 128 MHz reference<br />

signal in the selected measurement window. The level of the 128-MHz signals can be selected by<br />

DIAGnostic:SERVice:CSOource[:POWer] command.<br />

Example: "DIAG:SERV:INP CAL"<br />

’ Selects the internal reference signal.<br />

Characteristics: *RST value: RF<br />

SCPI: device-specific<br />

Mode: A<br />

If the rectangle calibration signal is switched on, the pulsed calibration signal is<br />

switched off.<br />

6.104 Operating Manual 1313.9681.12 - 01