R&S <strong>FSQ</strong> Instrument Functions<br />

LO/IF ports for external mixers - Option R&S FSU-B21<br />

The example of operation is described in the following steps:<br />

1. Test setup<br />

2. Activating the external mixing and selecting the mode<br />

3. Default setting<br />

4. Level correction<br />

5. Frequency-dependent level correction<br />

6. Level correction with average value<br />

7. Taking into account the cable loss in the IF path<br />

8. Functions for signal identification<br />

1. Test setup<br />

➢ Connect the 'LO OUT / IF IN' output of the R&S <strong>FSQ</strong> to the LO/IF port of the<br />

external mixer.<br />

➢ Connect the multiplier to the external mixer.<br />

➢ Apply a sinewave signal with f = 14.5 GHz to the input of the multiplier.<br />

2. Activating the external mixing and selecting the mode<br />

➢ Activate the external mixing with<br />

[ FREQUENCY : EXTERNAL MIXER ]<br />

3. Default setting<br />

➢ Prior to the measurement select the required band (in this case the V band)<br />

with<br />

[ FREQUENCY : EXTERNAL MIXER : SELECT BAND]<br />

➢ Select the V band using the cursor keys and ENTER<br />

4. Level correction<br />

The conversion loss of the mixer can be taken into account both as a function of<br />

frequency and with an average value. The frequency-dependent level correction<br />

is used in this example for obtaining a higher precision. First select a table valid<br />

for the selected band. As an alternative, level correction based on average value<br />

can be used (see “Level correction with average value” on page 4.300).<br />

5. Frequency-dependent level correction<br />

➢ Activate a selection list which contains all tables stored on the hard disk of the<br />

R&S <strong>FSQ</strong> with<br />

[FREQUENCY : EXTERNAL MIXER : SELECT BAND]<br />

➢ Select the CONV LOSS TABLE field by means of the cursor keys.<br />

➢ If a file with correction data already exists for the mixer used, select a table<br />

valid with the cursor keys or the rotary knob and confirm with ENTER.<br />

4.298 Operating Manual 1313.9681.12 - 01