R&S <strong>FSQ</strong> Instrument Functions<br />

Accuracy Enhancement<br />

label<br />

Medium APX<br />

(approximation)<br />

- Aborted<br />

normalization<br />

Tracking Generator – Option R&S FSU-B9<br />

Reason/Limitation<br />

Change of the following settings:<br />

• coupling (RBW, VBW, SWT)<br />

• reference level, RF attenuation<br />

• start or stop frequency<br />

• output level of tracking generator<br />

• frequency offset of tracking generator<br />

• detector (max. peak, min. peak, sample, etc.)<br />

Change of frequency:<br />

• max. 625 points within the set sweep limits<br />

(corresponds to a doubling of the span)<br />

• 625 and more extrapolated points within the current<br />

sweep limits (in case of span doubling)<br />

At a reference level (REF LEVEL) of -10 dBm and at a tracking generator output<br />

level of the same value the R&S <strong>FSQ</strong> operates without overrange reserve, i.e. the<br />

R&S <strong>FSQ</strong> is in danger of being overloaded if a signal is applied whose amplitude<br />

is higher than the reference line. In this case, either the message OVLD for overload<br />

is displayed in the status line or the display range is exceeded (clipping of the<br />

trace at the upper diagram border = Overrange).<br />

Overloading can be avoided as follows:<br />

• Reducing the output level of the tracking generator (SOURCE POWER,<br />

NETWORK menu)<br />

• Increasing the reference level (REF LEVEL, AMPT menu)<br />

4.230 Operating Manual 1313.9681.12 - 01