R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

SYSTem:DISPlay:FPANel ON | OFF<br />

SYSTem Subsystem<br />

This command activates or deactivates the display of the front panel keys on the screen.<br />

With the display activated, the instrument can be operated on the screen using the mouse by pressing<br />

the corresponding buttons. This may be useful if the instrument is operated in a detached station by<br />

means of a remote program such as PCANYWHERE.<br />

Example: "SYST:DISP:FPAN ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:DISPlay:UPDate ON | OFF<br />

This command switches on or off the update of all display elements during remote control.<br />

Example: "SYST:DISP:UPD ON"<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:ERRor?<br />

This command queries the earliest entry in the error queue, and deletes it after the readout.<br />

Positive error numbers indicate device-specific errors, negative error numbers are error messages<br />

defined by SCPI (cf. chapter “Error Messages”). If the error queue is empty, the error number 0, "no<br />

error", is returned. This command is identical with the command STATus:QUEue:NEXT?.<br />

This command is a query and therefore has no *RST value.<br />

Example: "SYST:ERR?"<br />

Characteristics: *RST value: –<br />

SCPI: conform<br />

Mode: A<br />

With the display of the front panel keys activated, the screen resolution of the unit<br />

is set to 1024x768. Thus, only a section of the whole screen is visible on the internal<br />

LCD display, which will be moved by mouse moves.<br />

For a full display of the user interface, an external monitor has to be connected to<br />

the rear panel.<br />

When the front panel display is deactivated, the original screen resolution is<br />

restored.<br />

The best performance is obtained when the display output is switched off during<br />

remote control.<br />