R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:LIMit:LOWer:OFFSet <br />

CALCulate Subsystem<br />

This command defines an offset for the y-axis of the selected relative lower limit line. In contrast to<br />

CALC:LIM:LOW:SHIFt, the line is not shifted by modifying the individual Y values but by means of<br />

an additive offset. The offset is independent of the measurement window.<br />

Example: "CALC:LIM2:LOW:OFFS 3dB"<br />

'Shifts limit line 2 in the corresponding measurement windows by 3 dB upwards.<br />

Characteristics: *RST value: 0<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:LOWer:SHIFt <br />

This command shifts a limit line by the indicated value in Y direction. In contrast to CALC:LIM:LOW:<br />

OFFS, the line is shifted by modifying the individual Y values but not by means of an additive offset.<br />

Up to 8 limit lines can be defined at the same time. The shift is independent of the measurement<br />

window.<br />

Example: "CALC:LIM3:LOW:SHIF 20DB"<br />

'Shifts all Y values of limit line 3 by 20 dB.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:LIMit:LOWer:SPACing LINear | LOGarithmic<br />

This command selects linear or logarithmic interpolation for the lower limit line.<br />

Example: "CALC:LIM:LOW:SPAC LIN"<br />

Characteristics: *RST value: LIN<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:LOWer:STATe ON | OFF<br />

This command switches on or off the indicated limit line. The limit check is activated separately with<br />

CALC:LIM:STAT ON.<br />

Up to 8 limit lines can be defined at the same time. The numeric suffixes of CALCulate indicate<br />

the measurement window.<br />

In analyzer mode, the result of the limit check can be queried with CALCulate:LIMit:<br />

FAIL?.<br />

Example: "CALC:LIM4:LOW:STAT ON"<br />

'Switches on limit line 4 (lower limit) in screen A.<br />

"CALC2:LIM4:LOW:STAT ON"<br />

'Switches on limit line 4 (lower limit) also in screen B.<br />

Characteristics: *RST value: OFF<br />

SCPI: conform<br />

Mode: A<br />

6.40 Operating Manual 1313.9681.12 - 01