R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E<br />

REMOTE state, the R&S <strong>FSQ</strong> is reconfigured such that the number of test points (trace points) corresponds<br />

to that of the 85xx families (1001 for 8566A/B and 8568A/B, 601 for 8560E to 8565E, 401 for<br />

8594E).<br />

6.23.9 Trace Data Output Formats<br />

All formats are supported for trace data output: display units (command O1), display units in two byte<br />

binary data (command O2 or TDF B and MDS W), display units in one byte binary data (command O4 or<br />

TDF B and MDS B) and physical values (commands O3 or TDF P). With format "display units" the level<br />

data will be converted into value range and resolution of the 8566/8568 models. On transition to<br />

REMOTE state the number of trace points will be reconfigured in order to be conform to the selected<br />

instrument model (1001 for 8566A/B and 8568 A/B, 601 for 8560E to 8565E, 401 for 8594E).<br />

6.23.10 Trace Data Input Formats<br />

Trace data input is only supported for binary date (TDF B, TDF A, TDF I, MDS W, MDS B).<br />

6.23.11 GPIB Status Reporting<br />

The assignment of status bits by commands R1, R2, R3, R4, RQS is supported starting at firmware version<br />

1.8x.<br />

The STB command and the serial poll respond with an 8-bit value having the following assignment:<br />

Bit enabled by RQS<br />

0 not used (value 0)<br />

1 Units key pressed<br />

2 End of Sweep<br />

3 Device Error<br />

4 Command Complete<br />

5 Illegal Command<br />

6 Service Request<br />

7 not used (value 0)<br />

Bits 0 and 7 are not used and always have the value 0.<br />

Please note that the R&S <strong>FSQ</strong> reports any key pressed on the front panel rather than only the unit keys<br />

if bit 1 was enabled.<br />

Another difference is the behavior of bit 6 when using the STB? query. On the HP analyzers this bit monitors<br />

the state of the SRQ line on the bus. On the R&S <strong>FSQ</strong> this is not possible. Therefore this bit is set,<br />

as soon as one of the bits 1 to 5 is set. It won't be reset by performing a serial poll.<br />

6.282 Operating Manual 1313.9681.12 - 01