R&S <strong>FSQ</strong> Instrument Functions<br />

External Generator Control – Option R&S FSP-B10<br />

Fig. 4.31 Normalized measurement, shifted with REF VALUE POSITION 50%<br />

Remote command: DISP:WIND:TRAC:Y:RPOS 10PCT<br />

REF VALUE The REF VALUE softkey activates the input of a level difference which is assigned<br />

to the reference line.<br />

In the default setting, the reference line corresponds to a level difference of 0 dB. If<br />

e.g. a 10-dB attenuator pad is inserted between DUT and R&S <strong>FSQ</strong> input between<br />

recording of the calibration data and normalization, the trace will be shifted down by<br />

10 dB. By entering a REF VALUE of –10 dB the reference line for difference formation<br />

can also be shifted down by 10 dB so that it will again coincide with the trace<br />

(see Fig. 4.32).<br />

4.240 Operating Manual 1313.9681.12 - 01