18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Instrument Functions<br />

SAVE LIMIT<br />

LINE<br />

Basic Settings<br />

In this manner, a new limit line can be easily generated based upon an existing limit<br />

line which has been shifted vertically and stored (SAVE LIMIT LINE softkey) under a<br />

new name (NAME softkey).<br />

Remote command: CALC:LIM3:CONT:UPP:SHIF 20dB<br />

CALC:LIM3:CONT:LOW:SHIF 20dB<br />

The SAVE LIMIT LINE softkey stores the currently edited limit line. The name can be<br />

entered in an input window (max. 8 characters)<br />

Remote command: --<br />

4.6.1.3 Display Lines<br />

Display lines help to evaluate a trace – as do markers. The function of a display line<br />

is comparable to that of a ruler that can be shifted on the trace in order to mark<br />

absolute values.<br />

The R&S <strong>FSQ</strong> provides two different types of display lines:<br />

• two horizontal level lines for marking levels – Display Line 1/2,<br />

• two vertical frequency or time lines for marking frequencies or points in time –<br />

Frequency/Time Line 1/2.<br />

Each line is identified by one of the following abbreviations:<br />

D1 Display Line 1<br />

D2 Display Line 2<br />

F1 Frequency Line 1<br />

F2 Frequency Line 2<br />

T1 Time Line 1<br />

T2 Time Line 2<br />

The level lines are continuous horizontal lines across the entire width of a diagram<br />

and can be shifted in y direction.<br />

The frequency or time lines are continuous vertical lines across the entire height of<br />

the diagram and can be shifted in x direction.<br />

The DISPLAY LINES submenu for activating and setting the display lines appears<br />

different depending on the display mode set in the active measurement window (frequency<br />

or time domain).<br />

If the spectrum is shown (span ≠ 0) the TIME LINE 1 and TIME LINE 2 softkeys are<br />

disabled, whereas in the time domain (span = 0) the FREQUENCY LINE 1 and<br />

FREQUENCY LINE 2 softkeys are not available.<br />

Working with display lines<br />

The softkeys for setting and switching the display lines on/off work like triple<br />

switches:<br />

Initial situation: The line is off (softkey with gray background)<br />

1st press: The line is switched on (softkey with red background) and the data input<br />

function is activated. The position of the display line can be selected by means of<br />

the rotary knob, the step keys or a numerical entry in the appropriate field. The data<br />

input function is disabled if another function is activated. The line, however, remains<br />

switched on (softkey with green background).<br />