R&S <strong>FSQ</strong> Instrument Functions<br />

• TYPE<br />

External Generator Control – Option R&S FSP-B10<br />

The field opens the list with the available generators:<br />

After completion of the selection, the remaining fields of the table are filled with<br />

the generator characteristics.<br />

A list of generator types supported by the R&S <strong>FSQ</strong> is to be found in section “List<br />

of Generator Types Supported by the R&S <strong>FSQ</strong>” on page 4.248.<br />

• IFC<br />

This field selects the interface type of external generator 1 or 2. The following<br />

types are available:<br />

– GPIB: GPIB only, suitable for all generators of other manufacturers and some<br />

<strong>Rohde</strong> & <strong>Schwarz</strong> instruments<br />

– TTL: GPIB and TTL interface for synchronization, for most of the <strong>Rohde</strong> &<br />

<strong>Schwarz</strong> generators, see table above.<br />

The two operating modes differ in the speed of the control: Whereas, with pure<br />

GPIB operation, each frequency to be set must be individually transferred to the<br />

generator, additional use of the TTL interface permits to program a total frequency<br />

list at once and subsequently perform the frequency stepping via TTL handshake,<br />

which is a big advantage in terms of speed.<br />

Generators equipped with the TTL interface can also be operated with GPIB only.<br />

Only one of the two generators can be operated with TTL interface at a time. The<br />

other generator must be configured for GPIB.<br />

• GPIB ADDR<br />

GPIB address of the respective generator. Addresses from 0 to 30 are possible.<br />

• MODE<br />

Operating mode of generator. The generator activated using the FREQUENCY<br />

SWEEP softkey is automatically set to remote mode (REMOTE), the other to<br />

manual operation (LOCAL).<br />

• FMIN F MAX<br />

Frequency range of generator. Select the start and stop frequency of the<br />

R&S <strong>FSQ</strong> in a way that the specified range is not exceeded. If the start frequency<br />

lies below F MIN, the generator is only switched on when F MIN is reached. If the<br />

stop frequency lies above F MAX, it is limited to F MAX when the generator is<br />

switched on using the EXT SRC ON/OFF softkey.<br />

• PMIN P MAX<br />

Level range of generator. This field defines the allowed input range for the<br />

POWER column in the FREQUENCY SWEEP table.<br />

Remote command: SYST:COMM:RDEV:GEN2:TYPE 'SMA01A'<br />

SYST:COMM:RDEV:GEN:LINK TTL<br />

SYST:COMM:GPIB:RDEV:GEN1:ADDR 28<br />