R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.6.7 CALCulate:MARKer:FUNCtion:SUMMary Subsystem<br />

CALCulate Subsystem<br />

This subsystem contains the commands for controlling the time domain power functions. These are provided<br />

in the marker subsystem for reasons of compatibility with the FSE family.<br />

CALCulate:MARKer:FUNCtion:SUMMary:AOFF<br />

This command switches off all time domain measurements in the selected measurement window. The<br />

function is independent of the marker selection, i.e. the numeric suffix of :MARKer is irrelevant.<br />

It is only available in the time domain (span = 0).<br />

Example: "CALC2:MARK:FUNC:SUMM:AOFF"<br />

'Switches off the time domain power measurement functions in screen B.<br />

Characteristics: *RST value: _<br />

SCPI: device-specific<br />

Mode: A-T<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:MARKer:FUNCtion:SUMMary:AVERage ON | OFF<br />

This command switches on or off averaging for the active time domain power measurement in the<br />

indicated window. The function is independent of the marker selection, i.e. the numeric suffix <br />

of :MARKer is irrelevant. It is only available in the time domain (span = 0).<br />

Averaging is reset by switching it off and on again.<br />

The number of results required for the calculation of average is defined with [SENSe:<br />

]AVERage:COUNt.<br />

It should be noted that synchronization to the end of averaging is only possible in single-sweep mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC2:MARK:FUNC:SUMM:AVER ON"<br />

'Switches on the calculation of average in screen B.<br />

"AVER:COUN 200"<br />

'Sets the measurement counter to 200.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A-T<br />

CALCulate:MARKer:FUNCtion:SUMMary:MEAN:AVERage:RESult?<br />

This command queries the result of the measurement of the averaged mean value in the selected<br />

measurement window. The query is only possible if averaging has been activated previously using<br />

CALCulate:MARKer:FUNCtion:SUMMary:AVERage.<br />

The function is independent of the marker selection, i.e. the numeric suffix of :MARKer is<br />

irrelevant. It is only available in the time domain (span = 0).<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

Operating Manual 1313.9681.12 - 01 6.81