R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.16.4 SENSe:DETector Subsystem<br />

SENSe Subsystem<br />

The SENSe:DETector subsystem controls the acquisition of measurement data via the selection of the<br />

detector for the corresponding trace. The measurement windows are selected by SENSe1 (screen A)<br />

and SENSe2 (screen B).<br />

[SENSe:]DETector[:FUNCtion] APEak | NEGative | POSitive | SAMPle | RMS | AVERage<br />

| QPEak<br />

This command selects the detector for the data acquisition of the selected trace and measurement<br />

window.<br />

• The APEak detector (AutoPeak) displays the positive and also the negative peak value of the noise<br />

floor. If a signal is detected, only the positive peak value is displayed.<br />

• The POSitive or NEGative detector only displays the positive or the negative peak value.<br />

• With the SAMPle detector the value measured at the sampling time is displayed.<br />

• The RMS detector evaluates the root mean square at each sweep point.<br />

• The AVERage detector evaluates the linear average power at each sweep point.<br />

• The QPEak detector performs a signal evaluation for EMC measurements.<br />

If QPEak is selected, the video filter is automatically switched off. The couplings between span and<br />

RBW as well as between RBW and sweep time are also switched off and restored on selecting another<br />

detector. A long sweep time should be selected so that the quasipeak detector can fully settle at each<br />

measurement point.<br />

The numeric suffix at DETector selects the trace to apply the detector to.<br />

Example: "DET POS"<br />

'Sets the detector in screen A to "positive peak".<br />

Characteristics: *RST value: APEak<br />

SCPI: conform<br />

Mode: A<br />

[SENSe:]DETector[:FUNCtion]:AUTO ON | OFF<br />

This command either couples the detector in the selected measurement window to the current trace<br />

setting or turns coupling off. The trace is selected by the numeric suffix at DETector.<br />

Example: "DET:AUTO OFF"<br />

Characteristics: *RST value: ON<br />

SCPI: conform<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.161