R&S <strong>FSQ</strong> Instrument Functions<br />

OCCUPIED<br />

BANDWIDTH<br />

OCCUP BW<br />

(ON OFF)<br />

% POWER<br />

BANDWIDTH<br />

CHANNEL<br />

BANDWIDTH<br />

<strong>Analyzer</strong> Mode<br />

Measurement of Occupied Bandwidth<br />

An important characteristics of a modulated signal is its occupied bandwidth. In a<br />

radio communications system for instance the occupied bandwidth must be limited<br />

to enable distortion-free transmission in adjacent channels. The occupied bandwidth<br />

is defined as the bandwidth containing a defined percentage of the total transmitted<br />

power. A percentage between 10% and 99.9% can be set on the R&S <strong>FSQ</strong>.<br />

OCCUP BW (ON OFF)<br />

% POWER BANDWIDTH<br />

CHANNEL BANDWIDTH<br />

ADJUST REF LVL<br />

ADJUST SETTINGS<br />

The OCCUPIED BANDWIDTH softkey activates measurement of the occupied<br />

bandwidth according to the current configuration and opens the submenu for configuring<br />

the measurement. The softkey is available only in frequency domain (span ><br />

0) and is highlighted when the measurement is switched on.<br />

In the spectrum display mode, this measurement determines the bandwidth that<br />

contains a predefined percentage of the power of the displayed frequency range (%<br />

POWER BANDWIDTH softkey). The occupied bandwidth is output in the marker display<br />

field and marked on the trace by temporary markers.<br />

• The softkey is only available in the frequency domain (span > 0).<br />

• The measurement is performed on the trace with marker 1. In order to evaluate<br />

another trace, marker 1 must be placed on another trace by means of SELECT<br />

TRACE in the MARKER menu<br />

The OCCUP BW ON/OFF softkey switches measurement of the occupied bandwidth<br />

on or off.<br />

Remote command: CALC:MARK:FUNC:POW:SEL OBW<br />

CALC:MARK:FUNC:POW:RES? OBW<br />

CALC:MARK:FUNC:POW OFF<br />

The % POWER BANDWIDTH softkey opens the entry of the percentage of power<br />

related to the total power in the displayed frequency range which defines the occupied<br />

bandwidth (percentage of total power).<br />

The valid range of values is 10% to 99.9%.<br />

Remote command: SENS:POW:BWID 99PCT<br />

The CHANNEL BANDWIDTH softkey opens an input window for defining the channel<br />

bandwidth for the transmission channel. For measurements in line with a specific<br />

transmission standard, the bandwidth specified by the standard for the transmission<br />

channel must be entered. The default setting is 14 kHz.<br />

The specified channel bandwidth is used for optimization of the test parameters of<br />

the R&S <strong>FSQ</strong> with ADJUST SETTINGS.<br />

Remote command: SENS:POW:ACH:BWID 14kHz<br />