R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:DELTamarker:FUNCtion:PNOise:AUTO ON | OFF<br />

CALCulate Subsystem<br />

This command adds an automatic peak search action for the reference fixed marker 1 at the end of<br />

each particular sweep. This function may be used for tracking of a drifting source whilst phase noise<br />

measurement. The delta marker 2 which shows the phase noise measurement result keeps the delta<br />

frequency value. Therefore the phase noise measurement in a certain offset is valid although the<br />

source is drifting. Only when the marker 2 is reaching the border of the span the delta marker value is<br />

adjusted to be within the span. Choose a larger span in such situations.<br />

Example: "CALC:DELT:FUNC:PNO 1"<br />

’Switches the phase noise measurement on.<br />

“CALC:DELT:FUNC:PNO:AUTO ON”<br />

’Activates the automatic peak search.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

The suffix at DELTamarker is ignored.<br />

CALCulate:DELTamarker:FUNCtion:PNOise:RESult?<br />

This command queries the result of the phase noise measurement in the selected measurement<br />

window. The measurement will be switched on, if necessary.<br />

Example: "CALC:DELT:FUNC:PNO:RES?"<br />

'Outputs the result of phase noise measurement of the selected delta marker in<br />

screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only a query and therefore has no *RST value.<br />

CALCulate:DELTamarker:FUNCtion:PNOise[:STATe] ON | OFF<br />

This command switches on or off the phase noise measurement with all active delta markers in the<br />

selected measurement window. The correction values for the bandwidth and the log amplifier are taken<br />

into account in the measurement.<br />

Marker 1 will be activated, if necessary, and a peak search will be performed. If marker 1 is activated,<br />

its position becomes the reference point for the measurement.<br />

The reference point can then be modified with commands CALCulate:DELTamarker:FUNCtion:<br />

FIXed:RPOint:X and ...:RPOint:Y independently of the position of marker 1 and of a trace (the<br />

same commands used for the measurement with fixed reference point).<br />

The numeric suffix with DELTamarker is not relevant for this command.<br />

Example: "CALC:DELT:FUNC:PNO ON"<br />

'Switches on the phase noise measurement with all delta markers in screen A.<br />

"CALC:DELT:FUNC:FIX:RPO:X 128 MHZ"<br />

'Sets the reference frequency to 128 MHz.<br />

"CALC:DELT:FUNC:FIX:RPO:Y 30 DBM"<br />

'Sets the reference level to +30 dBm<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

6.16 Operating Manual 1313.9681.12 - 01