R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

SENSe Subsystem<br />

1. Setting up the instrument and at the same time querying the result list:<br />

This method ensures the smallest delay between measurement and the output of the measured<br />

values, but requires the control computer to wait actively for the response of the instrument.<br />

2. Setting up the instrument and querying the result list after synchronization to the end of measurement:<br />

With this method the control computer can be used for other activities while the instrument is<br />

performing the measurement at the cost of additional time needed for synchronization via service<br />

request.<br />

[SENSe:]MPOWer:RESult[:LIST]?<br />

This command queries the results of a multiple burst power measurement as configured and initiated<br />

with SENSe:MPOWer[:SEQuence]. The results are output in a comma-separated list of floating point<br />

values. The unit used for the return values is always dBm.<br />

This command may be used to obtain measurement results in an asynchronous way using the service<br />

request mechanism for synchronization with the end of the measurement.<br />

If no measurement results are available, the command will return a query error.<br />

Example: *ESE 1<br />

*SRE 32<br />

'Configuration of status reporting systems for the generation of an SRQ on<br />

operation complete<br />

SENSe:MPOWer<br />

935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC<br />

'Configuring and starting the measurement<br />

...<br />

'Further actions of the control computer during measurement<br />

On SRQ:<br />

'Response to service request<br />

SENSe:MPOWer:RESult?<br />

Characteristics: *RST value: --<br />

SCPI: instrument-specific<br />

Mode: A-F, A-T<br />

[SENSe:]MPOWer:RESult:MIN?<br />

This command queries the minimum power value in a multiple burst power measurement as<br />

configured and initiated with SENSe:MPOWer[:SEQuence]. The unit used for the return values is<br />

always dBm.<br />

If no measurement result is available, the command will return a query error.<br />

Example: *ESE 1<br />

*SRE 32<br />

'Configuration of status reporting systems for the generation of an SRQ on<br />

operation complete<br />

SENSe:MPOWer<br />

935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC<br />

'Configuring and starting the measurement<br />

...<br />

'Further actions of the control computer during measurement<br />

Operating Manual 1313.9681.12 - 01 6.191