R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]CORRection:CVL:CATalog?<br />

This command queries all the names of the conversion loss tables saved on the hard disc.<br />

The syntax of the output is:<br />

,,<br />

,,,,...,,<br />

,<br />

Example: ":CORR:CVL:CAT?"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only available if option R&S FSU-B21 is installed.<br />

[SENSe:]CORRection:CVL:CLEar<br />

SENSe Subsystem<br />

This command deletes the selected Conversion Loss Table. Prior to this command, the command<br />

SENS:CORR:CVL:SEL must be sent.<br />

Example: "CORR:CVL:SEL 'LOSS_TAB'"<br />

"CORR:CVL:CLE"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event which is why it is not assigned an *RST value.<br />

This command is only available with the option R&S FSU-B21.<br />

[SENSe:]CORRection:CVL:COMMent <br />

This command defines the comment in the Conversion Loss Table. Prior to this command, the<br />

command SENS:CORR:CVL:SEL must be sent.<br />

Parameter: ::= Comment of mixer with a maximum of 60 characters<br />

Example: "CORR:CVL:SEL 'LOSS_TAB'"<br />

"CORR:CVL:COMMENT 'MIXER FOR BAND U'"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only available with the option R&S FSU-B21.<br />

[SENSe:]CORRection:CVL:DATA ,..<br />

This command defines the reference values of the selected Conversion Loss Tables. The values are<br />

entered as a result of frequency/level pairs. The frequencies have to be sent in ascending order. A<br />

maximum of 50 pairs of frequency/level may be entered. Prior to this command, the command SENS:<br />

CORR:CVL:SEL must be sent.<br />

Example: "CORR:CVL:SEL 'LOSS_TAB'"<br />

"CORR:CVL:DATA 1MHZ,-30DB,2MHZ,-40DB"<br />

6.154 Operating Manual 1313.9681.12 - 01