18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

R&S <strong>FSQ</strong> Instrument Functions<br />

YIG FILTER<br />

(ON OFF)<br />

IQ DITHER ON<br />

OFF<br />

Basic Settings<br />

In order to carry out broadband signal analysis, the YIG filter at the input of the<br />

R&S <strong>FSQ</strong> can be removed from the signal path.<br />

The YIG (ON OFF) softkey switches the YIG to the RF path by means of relays or<br />

bypasses the filter. With bypassed YIG filters, the maximum bandwidth can be used<br />

for signal analysis. However, image-frequency rejection is no longer ensured.<br />

Remote command: INP:FILT:YIG ON|OFF<br />

If option R&S <strong>FSQ</strong>-B72 is installed, this softkey switches the dither signal on and off.<br />

The default setting in OFF.<br />

Remote command: IQ:DITH ON|OFF<br />

RF PATH The RF PATH softkey selects the RF Input Path of the analyzer. This softkey is not<br />

available in FFT analyzer mode.<br />

This softkey is only available with option R&S <strong>FSQ</strong>-B71 or R&S <strong>FSQ</strong>-B17. Refer to<br />

the manual of the R&S <strong>FSQ</strong>-B71 or the R&S <strong>FSQ</strong>-B17 for details.<br />

BASEBAND<br />

ANALOG<br />

BASEBAND<br />

DIGITAL<br />

DIGITAL BB INFO<br />

EX-IQ BOX<br />

The YIG filter operates only if the frequency in the signal path exceeds 3.6 GHz.<br />

The command has, therefore, no effect if the frequency is below this threshold.<br />

Remote command: INP:SEL RF | AIQ | DIQ<br />

The BASEBAND ANALOG softkey opens a submenu to configure the analog baseband<br />

input.<br />

This softkey is only available with option R&S <strong>FSQ</strong>-B71. Refer to the R&S <strong>FSQ</strong>-B71<br />

manual for details on the functions available in the submenu.<br />

The BASEBAND DIGITAL softkey opens a submenu to configure the digital baseband<br />

input.<br />

This softkey is only available with option R&S <strong>FSQ</strong>-B17.<br />

DIGITAL IN FULL SCALE<br />

DIGITAL IN SAMPLE RATE<br />

FULL SCALE AUTO SET /<br />

SAMPLE RATE AUTO SET<br />

DIGITAL BB INFO<br />

EX-IQ BOX<br />

Operating Manual 1313.9681.12 - 01 4.175

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!