18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E<br />

6.23.7 Model-Dependent Default Settings<br />

If the GPIB language is switched over to an 85xx model, the GPIB address will automatically be switched<br />

over to 18 provided that the default address of the R&S <strong>FSQ</strong> (20) is still set. If a different value is set, this<br />

value will be maintained. Upon return to SCPI, this address will remain unchanged.<br />

The following table shows the default settings obtained after a change of the GPIB language and for the<br />

commands IP, KST and RESET:<br />

Model # of Trace Points Start Freq. Stop Freq. Ref Level Input Coupling<br />

8566A/B 1001 2 GHz 22 GHz 0 dBm DC<br />

8568A/B 1001 0 Hz 1.5 GHz 0 dBm AC<br />

8560E 601 0 Hz 2.9 GHz 0 dBm AC<br />

8561E 601 0 Hz 6.5 GHz 0 dBm AC<br />

8562E 601 0 Hz 13.2 GHz 0 dBm AC<br />

8563E 601 0 Hz 26.5 GHz 0 dBm AC<br />

8564E 601 0 Hz 40 GHz 0 dBm AC<br />

8565E 601 0 Hz 50 GHz 0 dBm AC<br />

8594E 401 0 Hz 3 GHz 0 dBm AC<br />

Notes regarding the set stop frequency:<br />

• The stop frequency given in the table may be limited to the corresponding<br />

frequency range of the R&S FSUP.<br />

• Command LF sets the stop frequency for 8566A/B to a maximum value of 2<br />

GHz.<br />

Note regarding the number of test points (trace points):<br />

• The number of trace points is switched over only upon transition to the REMOTE<br />

state.<br />

6.23.8 Data Output Formats<br />

In the case of the SCPI and IEEE488.2 standards, the output formats for numerical data are flexible to a<br />

large extent. The output format for the HP units, by contrast, is accurately defined with respect to the<br />

number of digits. The memory areas for reading instrument data have therefore been adapted accordingly<br />

in the remote-control programs for instruments of this series.<br />

Therefore, in response to a query, the R&S <strong>FSQ</strong> returns data of the same structure as that used by the<br />

original instruments; this applies in particular to the number of characters returned.<br />

Two formats are currently supported when trace data is output: Display Units (command O1) and physical<br />

values (command O2, O3 or TDF P). As to the "Display Units" format, the level data of the R&S <strong>FSQ</strong><br />

is converted to match the value range and the resolution of the 8566/8568 series. Upon transition to the<br />

Operating Manual 1313.9681.12 - 01 6.281

Hooray! Your file is uploaded and ready to be published.

Saved successfully!