R&S <strong>FSQ</strong> Instrument Functions<br />

PERCENT<br />

MARKER<br />

<strong>Analyzer</strong> Mode<br />

If the CCDF function is active, the PERCENT MARKER softkey allows to position<br />

marker 1 by entering a probability value. Thus, the power which is exceeded with a<br />

given probability can be determined very easily.<br />

If marker 1 is in the switched-off state, it will be switched on automatically.<br />

Remote command: CALC:MARK:Y:PERC 0...100%<br />

RES BW The RES BW softkey sets the resolution bandwidth in the menu STATISTIC FUNC-<br />

TION directly without switching to the corresponding menu (BW). The function of the<br />

softkey is identical to the softkey RES BW MANUAL in the BW menu.<br />

For correct measurement of the signal statistics the resolution bandwidth has to be<br />

wider than the signal bandwidth in order to transmit the actual peaks of the signal<br />

amplitude correctly. Video bandwidth is set to 10 MHz automatically with a statistic<br />

function switched on.<br />

NO OF<br />

SAMPLES<br />

Remote command: BAND 3 MHz<br />

The NO OF SAMPLES softkey sets the number of power measurements taken into<br />

account for the statistics.<br />

Please note that the overall measurement time is influenced by the number of samples<br />

selected as well as by the resolution bandwidth set up for the measurement as<br />

the resolution bandwidth directly influences the sampling rate.<br />

Remote command: CALC:STAT:NSAM <br />

SCALING The SCALING softkey opens a submenu that allows changing the scaling parameters<br />

for both the x- and the y-axis.<br />

X-AXIS REF<br />

LEVEL<br />

X-AXIS REF LEVEL<br />

X-AXIS RANGE<br />

Y-UNIT (% ABS)<br />

Y-AXIS MAX VALUE<br />

Y-AXIS MIN VALUE<br />

ADJUST SETTINGS<br />

DEFAULT SETTINGS<br />

MEAN PWR POSITION<br />

The X-AXIS REF LEVEL softkey changes the level settings of the instrument and<br />

sets the maximum power to be measured.<br />

The function is identical to softkey REF LEVEL in AMPT menu.<br />

For the APD function this value is mapped to the right diagram border. For the<br />

CCDF function there is no direct representation of this value on the diagram as the<br />

x-axis is scaled relatively to the MEAN POWER measured.<br />

Remote command: CALC:STAT:SCAL:X:RLEV <br />