R&S <strong>FSQ</strong> Instrument Functions<br />

• Max Transfer Rate: Maximum interface clock rate to transfer the I/Q data.<br />

Basic Settings<br />

• Connection Protocol: State of the connection protocol. The analyzer is able to<br />

communicate with the sending or receiving device.<br />

• PRBS Test Deskewing: An alignment process is started if the output or input of the<br />

R&S <strong>FSQ</strong>-B17 is connected to a digital baseband source/sink. The current state<br />

of this process is shown here. Possible states are "not yet started", "failed" or<br />

"passed".<br />

This softkey is only available with option R&S <strong>FSQ</strong>-B17.<br />

Remote command: INP:DIQ:CDEV?<br />

OUTP:DIQ:CDEV?<br />

EX-IQ BOX The EX-IQ BOX softkey opens a dialog to configure an R&S EX-IQ-Box connected<br />

to the digital baseband input or output.<br />

This softkey is only available with option R&S <strong>FSQ</strong>-B17.<br />

For more information refer to the manual for option R&S <strong>FSQ</strong>-B17.<br />

4.6.3.4 RF Preamplifier<br />

To improve the noise figure, a low-noise preamplifier with variable gain at the RF<br />

input can be switched into the signal path.<br />

PREAMP The PREAMP softkey switches the preamplifier on or off and activates the entry of<br />

the preamplifier gain. The preamplifier is switched off by pressing the softkey again.<br />

The only possible value with option el. attenuator (B25) is 20 dB.<br />

Remote command: INP:GAIN:STAT ON<br />

'Switches the 20 dB preamplifier on.<br />

The PREAMP softkey is only available with option el. attenuator (B25) or option<br />

low noise preamp (B24).<br />

4.6.3.5 Transducer<br />

Activating Transducer Factors<br />

The TRANSDUCER softkey opens a submenu enabling the user to activate or deactivate<br />

defined transducer factors, to generate new transducer factors or to edit existing<br />

ones. A table with the transducer factors defined is displayed.<br />

As soon as a transducer is activated, the unit of the transducer is automatically used<br />

for all the level settings and outputs. The unit cannot be changed in the AMPT menu<br />

since the R&S <strong>FSQ</strong> and the transducer used are regarded as one measuring instrument.<br />

Only if the transducer has the unit dB, will the unit originally set on the<br />

R&S <strong>FSQ</strong> be maintained and can be changed.<br />

If a transducer factor is active, TDF is displayed in the enhancement labels column.<br />

After all transducers have been switched off, the R&S <strong>FSQ</strong> returns to the unit that<br />

was used before a transducer was activated.<br />

Operating Manual 1313.9681.12 - 01 4.177