R&S <strong>FSQ</strong> Instrument Functions<br />

RF ATTEN<br />

AUTO<br />

NOISE CORR<br />

(ON OFF)<br />

REF LEVEL<br />

POSITION<br />

REF LEVEL<br />

OFFSET<br />

<strong>Analyzer</strong> Mode<br />

The attenuation can be set in 5 dB steps between 0 and 75 dB.<br />

Other entries will be rounded to the next higher integer value.<br />

If the defined reference level cannot be set for the given RF attenuation, the reference<br />

level will be adjusted accordingly and the warning "Limit reached" will be output.<br />

The 0 dB value can be entered only via the numeric keypad in order to protect the<br />

input mixer against accidental overload.<br />

Remote command: INP:ATT 40 DB<br />

The RF ATTEN AUTO softkey sets the RF attenuation automatically as a function of<br />

the selected reference level.<br />

This ensures that the optimum RF attenuation desired by the user is always used.<br />

RF ATTEN AUTO is the default setting.<br />

Remote command: INP:ATT:AUTO ON<br />

If active, the R&S <strong>FSQ</strong> corrects the results by its inherent noise. Noise correction<br />

increases the dynamic range.<br />

After you activate noise correction, the R&S <strong>FSQ</strong> performs a reference measurement<br />

of its inherent noise. In the actual measurement, the R&S <strong>FSQ</strong> then substracts<br />

the noise power from the power in the channel that is measured.<br />

The inherent noise depends on the center frequency, resolution bandwidth and level<br />

setting. Therefore, the R&S <strong>FSQ</strong> deactivates noise correction if you change one<br />

these parameters. The R&S <strong>FSQ</strong> shows a message that noise correction is inactive.<br />

The R&S <strong>FSQ</strong> also deactivates noise correction after you select another measurement<br />

(e.g. channel power, spectrum emission mask etc.).<br />

After the R&S <strong>FSQ</strong> deactivates noise correction, you have to activate it again manually.<br />

The R&S <strong>FSQ</strong> performs a new reference measurement.<br />

Remote command: POW:NCOR ON<br />

The REF LEVEL POSITION softkey allows the reference level position to be<br />

entered.<br />

The setting range is from -200 to +200%, 0% corresponding to the lower and 100%<br />

to the upper limit of the diagram.<br />

Remote command: DISP:WIND:TRAC:RPOS 100PCT<br />

The REF LEVEL OFFSET softkey allows the arithmetic level offset to be entered.<br />

This offset is added to the measured level irrespective of the selected unit. The scaling<br />

of the y-axis is changed accordingly.<br />

The setting range is ±200 dB in 0.1 dB steps.<br />

Remote command: DISP:WIND:TRAC:RLEV:OFFS -10dB<br />

Operating Manual 1313.9681.12 - 01 4.17