R&S <strong>FSQ</strong> Instrument Functions<br />

GEN REF INT /<br />

EXT<br />

External Generator Control – Option R&S FSP-B10<br />

If the offset is adjusted in a way that the sweep of the generator crosses the 0 Hz<br />

Frequency, it is indicated by the additional statement “via 0 Hz”.<br />

Example for reverse sweep via 0 Hz<br />

F<strong>Analyzer</strong>Start= 100 MHz<br />

F<strong>Analyzer</strong>Stop = 200 MHz<br />

FOffset = -150 MHz<br />

Numerator = Denominator = 1<br />

→ FGeneratorStart = 50 MHz<br />

→ FGeneratorStop = 50 MHz via 0 Hz<br />

Remote command: :SOUR:EXT:FREQ:NUM 1<br />

:SOUR:EXT:FREQ:DEN 1<br />

:SOUR:EXT:FREQ:OFFS -300MHZ<br />

• RESULT<br />

The frequency range of the generator resulting from the calculation formula. An<br />

asterisk (*) after the upper limit indicates that the stop frequency of the R&S <strong>FSQ</strong><br />

must be adapted when the generator is switched on in order not to exceed its<br />

maximum frequency. In the following illustration, this is true for the upper<br />

generator at a stop frequency of 3.2 GHz of the R&S <strong>FSQ</strong>, whereas the lower<br />

generator does not yet require an adaptation:<br />

Remote command: SOUR:EXT:POW –30dBm<br />

SOUR:EXT:FREQ:NUM 4<br />

SOUR:EXT:FREQ:DEN 3<br />

SOUR:EXT:FREQ:OFFS 100MHZ<br />

The GEN REF INT / EXT softkey switches over the reference oscillator of the generator<br />

(switch over between internal and external reference source). Selection EXT<br />

allows connecting the external generator to an external reference frequency source.<br />

The internal reference source is selected as the default setting.<br />

Remote command: SOUR:EXT1:ROSC INT<br />

Operating Manual 1313.9681.12 - 01 4.251