R&S <strong>FSQ</strong> Instrument Functions<br />

SWEEPTIME<br />

MANUAL<br />

SWEEPTIME<br />

AUTO<br />

SWEEP<br />

COUNT<br />

SWEEP<br />

POINTS<br />

<strong>Analyzer</strong> Mode<br />

The SWEEPTIME MANUAL softkey activates the window for entering the sweep<br />

time manually (see also BW menu).<br />

Remote command: SWE:TIME 10s<br />

The SWEEPTIME AUTO softkey activates the automatic selection of the sweep time<br />

as a function of the bandwidth of the resolution and video filters (see also BW<br />

menu).<br />

Remote command: SWE:TIME:AUTO ON<br />

The SWEEP COUNT softkey activates the window for the entry of the number of<br />

sweeps to be performed by R&S <strong>FSQ</strong> after a single sweep has been started. If<br />

Trace Average, Max Hold or Min Hold is activated, this also determines the number<br />

of averaging or maximum search procedures.<br />

Example<br />

[TRACE1: MAX HOLD]<br />

[SWEEP: SWEEP COUNT: {10} ENTER]<br />

[SINGLE SWEEP]<br />

R&S <strong>FSQ</strong> performs the Max Hold function over 10 sweeps.<br />

The permissible range for the sweep count is 0 to 32767. For sweep count = 0 or 1,<br />

one sweep is performed. For trace averaging in the continuous-sweep mode,<br />

R&S <strong>FSQ</strong> performs running averaging over 10 sweeps if sweep count = 0; if sweep<br />

count = 1, no averaging, maxhold or minhold is performed.<br />

The sweep count is valid for all the traces in a diagram.<br />

The number of sweeps set in the TRACE menu is the same as that in the SWEEP<br />

menu.<br />

If SINGLE SWEEP is selected, the measurement stops after the selected number<br />

of sweeps has been performed.<br />

Remote command: SWE:COUN 64<br />

The SWEEP POINTS softkey selects the number of measurement samples<br />

acquired during a sweep.<br />

The following numbers of points per sweep are available: 155, 201, 301, 313, 401,<br />

501, 601, 625 (default), 701, ..., 30001<br />

An increment of 100 is possible for number of points ≥ 201.<br />

Additionally 1251 and 1999 points are available.<br />

The autopeak detector will be disabled while the number of points per sweep is set<br />

to another value than the default.<br />

Remote command: SWE:POIN 501<br />