R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

On SRQ:<br />

'Response to service request<br />

SENSe:MPOWer:RESult:MIN?<br />

Characteristics: *RST value: --<br />

SCPI: instrument-specific<br />

Mode: A-F, A-T<br />

SENSe Subsystem<br />

[SENSe:]MPOWer[:SEQuence] ,,,,<br />

,,,<br />

This command configures the instrument setup for multiple burst power measurement and starts a<br />

measurement sequence. When synchronizing the command with *OPC, a service request is<br />

generated as soon as the defined number of individual measurements (# of meas) is reached.<br />

To reduce the setting time, the setup is performed simultaneously for all selected parameters.<br />

The command in the form of a query makes the instrument settings, performs the defined number of<br />

measurements and outputs the measurement results list.<br />

Parameter: <br />

Receive frequency for the burst signals to be measured (= center frequency in<br />

manual operation) The range is 0 Hz to max. frequency, depending on instrument<br />

model<br />

<br />

Resolution bandwidth for the measurement<br />

The range is from 10 Hz to 20 MHz in steps of 1, 2 3, 5, 10 steps and, additionally<br />

a bandwidth of 50 MHz.<br />

<br />

Time span during which measurement samples are sampled for RMS / peak<br />

measurement The type of measurement is selected by . The range<br />

is from 1 us to 30 s.<br />

<br />

Possible settings of the trigger signal source are:<br />

EXTernal:<br />

The trigger signal is fed from the "Ext. Trigger/Gate" input on the rear of the unit.<br />

VIDeo:<br />

The internal video signal is used as trigger signal.<br />

IFPower:<br />

The internal IF signal with 10 MHz bandwidth is used as the trigger signal.<br />

<br />

<strong>Signal</strong> level at which the trigger becomes active.<br />

For = VIDeo this is the level of the video signal as a percentage<br />

of the diagram height.<br />

For = IFPower this is the signal level in dBm.<br />

If = EXTernal is selected, the value entered here is ignored, as in<br />

this case the trigger input uses the normal trigger settings made via the TRIG key.<br />

The range for trigger source video is 0 – 100PCT.<br />

6.192 Operating Manual 1313.9681.12 - 01