R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Mode: A<br />

SOURce:POWer:STARt -30 dBm to +5 dBm<br />

SOURce Subsystem<br />

This command defines the start power of the power sweep. The start value can be smaller than the<br />

start value.<br />

The start power can be set between -30 dBm and +5 dBm.<br />

With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.<br />

Example: "SOUR:POW:STAR 0dBm"<br />

'sets the end level to 0 dBm in screen A.<br />

Characteristics: *RST value: 0 dBm<br />

SCPI: conform<br />

Mode: A<br />

SOURce:POWer:STOP -30 dBm to +5 dBm<br />

This command defines the stop power of the power sweep. The stop power can be set between -30<br />

dBm and +5 dBm.<br />

With the option R&S FSU-B12 the power can be set between -100 and + 5 dBm.<br />

The stop value can be smaller than the start value.<br />

Example: "SOUR:POW:STOP 0dBm"<br />

'sets the end level to 0 dBm in screen A.<br />

Characteristics: *RST value: 0 dBm<br />

SCPI: conform<br />

Mode: A<br />

6.214 Operating Manual 1313.9681.12 - 01