R&S <strong>FSQ</strong> Getting Started<br />

Measurements on Modulated <strong>Signal</strong>s<br />

The R&S SMIQ has to be equipped with options R&S SMIQ-B10 or R&S SMIQ-B20<br />

(modulation coder) and R&S SMIQ-B11 (data generator).<br />

Option R&S SMIQ-Z5 is required to trigger the R&S <strong>FSQ</strong>. This option is connected<br />

to the R&S SMIQ’s parallel output port. The BNC output Trigger 1 of the R&S SMIQ-<br />

Z5 provides a TTL trigger signal on the rising edge of the IS136 burst, which is used<br />

to start the R&S <strong>FSQ</strong> sweep in the Gated Sweep mode.<br />

The R&S <strong>FSQ</strong>’s IF power trigger is not suitable for IS136. It triggers on every<br />

level edge of the input signal. Since the modulation of the IS136 signal causes<br />

level dips even during the transmit burst, there is no way of ensuring that the<br />

R&S <strong>FSQ</strong> is only triggered on the burst edge.<br />

Settings on signal generator R&S SMIQ<br />

Switch the signal generator to the IS136 burst mode (time slots 1 and 4 are switched<br />

on, the other time slots are switched off).<br />

The R&S SMIQ is set as follows to generate the signal:<br />

1. Press the PRESET key.<br />

2. Press the FREQ key and enter 850 MHz.<br />

3. Press the LEVEL key and enter -20 dBm.<br />

4. Press the RETURN key.<br />

5. Select DIGITAL STANDARD using the rotary knob and press the SELECT key.<br />

6. Select NADC using the rotary knob and press the SELECT key.<br />

7. Press the SELECT key.<br />

8. Select ON using the rotary knob and press the SELECT key.<br />

9. Press the RETURN key.<br />

10.Keep turning the rotary knob until SAVE/RECALL FRAME appears in the list and<br />

select the menu item SAVE/RECALL FRAME using the SELECT key.<br />

11.The cursor is set to GET PREDEFINED FRAME.<br />

12.Press the SELECT key.<br />

13.Select UP1TCH using the rotary knob and press the SELECT key.<br />

In the following operating sequence for the R&S <strong>FSQ</strong>, it is assumed that steps 1 to 6<br />

of 2.5.3.2“Measurement Example 2 – Measuring the Adjacent Channel Power of an<br />

IS136 TDMA <strong>Signal</strong>” on page 2.39 have already been performed.<br />

Configuring the Gated Sweep function on the R&S <strong>FSQ</strong><br />

➢ Press the TRIG key.<br />

➢ Press the GATED TRIGGER softkey.<br />

➢ Press the EXTERN softkey.<br />

➢ Press the GATE SETTINGS � softkey.<br />

The R&S <strong>FSQ</strong> switches to time domain measurement so that the setting of the<br />

Gated Sweep parameters can be checked visually.<br />

➢ Press the SWEEPTIME softkey and enter 10 ms.<br />

Exactly one TDMA burst will be displayed.<br />

Operating Manual 1313.9681.12 - 01 2.43