R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:MARKer:X 0 to MAX (frequency | sweep time)<br />

This command positions the selected marker to the indicated :<br />

• frequency (span > 0)<br />

• time (span = 0)<br />

• level (APD/CCDF measurement)<br />

If marker 2, 3 or 4 is selected and used as delta marker, it is switched to marker mode.<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

Example: "CALC1:MARK2:X 10.7MHz"<br />

'Positions marker 2 in screen A to frequency 10.7 MHz.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:X:SLIMits:LEFT 0 to MAX (frequency | sweep time)<br />

CALCulate Subsystem<br />

This command sets the left limit of the search range for markers and delta markers in the selected<br />

measurement window. Depending on the x-axis domain, the indicated value defines:<br />

• frequency (span > 0)<br />

• time (span = 0)<br />

The function is independent of the selection of a marker, i.e. the numeric suffix in MARKer is<br />

irrelevant.<br />

If the time domain power measurement is active, this command limits the evaluation range to the trace.<br />

Example: "CALC:MARK:X:SLIM ON"<br />

'Switches the search limit function on for screen A.<br />

"CALC:MARK:X:SLIM:LEFT 10MHz"<br />

'Sets the left limit of the search range in screen A to 10 MHz.<br />

Characteristics: *RST value: - (is set to the left diagram border on switching on search limits)<br />

SCPI: device-specific<br />

Mode: A<br />

The function is only available if the search limit for marker and delta marker is<br />

switched on (CALC:MARK:X:SLIM ON).<br />

CALCulate:MARKer:X:SLIMits:RIGHt 0 to MAX (frequency | sweep time)<br />

This command sets the right limit of the search range for markers and delta markers in the selected<br />

measurement window. Depending on the x-axis domain the indicated value defines a frequency (span<br />

> 0) or time (span = 0). The function is independent of the selection of a marker, i.e. the numeric suffix<br />

in MARKer is irrelevant.<br />

If the time domain power measurement is active, this command limits the evaluation range to the trace.<br />

Operating Manual 1313.9681.12 - 01 6.51