R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.10 CALCulate:THReshold Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:THReshold subsystem controls the threshold value for the maximum/minimum search of<br />

markers. The measurement windows are selected by CALCulate 1 (screen A) or 2 (screen B).<br />

CALCulate:THReshold MINimum to MAXimum (depending on current unit)<br />

This command defines the threshold value for the maximum/minimum search of markers with marker<br />

search functions MAX PEAK, NEXT PEAK, etc. in the selected measurement window. The associated<br />

display line is automatically switched on.<br />

Example: "CALC:THR -82DBM"<br />

'Sets the threshold value for screen A to -82 dBm.<br />

Characteristics: *RST value: - (STATe to OFF)<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:THReshold:STATe ON | OFF<br />

This command switches on or off the threshold line in the selected measurement window. The unit<br />

depends on the setting performed with CALC:UNIT.<br />

Example: "CALC2:THR:STAT ON"<br />

'Switches on the threshold line in screen B.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

6.98 Operating Manual 1313.9681.12 - 01