R&S <strong>FSQ</strong> Instrument Functions<br />

<strong>Analyzer</strong> Mode<br />

SWEEPTIME The SWEEPTIME softkey enables the user to change the time axis to obtain a<br />

higher resolution for positioning gate delay and gate length.<br />

When this is to be done, the sweep time temporarily changes; the original value is<br />

restored when the menu is quit.<br />

Remote command: --<br />

Measurement example<br />

The modulation spectrum of a GSM 900 signal is to be measured using the gatedsweep<br />

function. The signal is generated by a <strong>Signal</strong> Generator whose RF output is<br />

directly connected to the RF input of R&S <strong>FSQ</strong>.<br />

Settings on the <strong>Signal</strong> Generator:<br />

FREQ: 802 MHz<br />

Level: 0 dBm: Return<br />

Digital Mod: Select: GMSK: Select<br />

Source: Select: PRBS: Select: Return<br />

Level Attenuation: Select: 60 dB: Return<br />

The <strong>Signal</strong> Generator supplies a GMSK-modulated TDMA signal (GSM).<br />

Settings on R&S <strong>FSQ</strong>:<br />

Conventions:<br />

[KEY] Menu called by this key. All information between the brackets<br />

refers to this menu.<br />

{Number} Numeric value to be entered for the parameter on hand.<br />

SOFTKEY Softkey to be used for making a selection or entering a value.<br />

[PRESET]<br />

[FREQ: CENTER {802} MHz]<br />

[SPAN {3.6} MHz]<br />

[AMPT: REF LEVEL {0} dBm: RF ATTEN MANUAL: {10} dB]<br />

[BW: RES BW MANUAL: {30} kHz]<br />

[TRACE: TRACE 1 DETECTOR: RMS]<br />

[SWEEP: SWEEPTIME MANUAL: {50} ms]<br />

[TRIG: EXTERN<br />

GATED TRIGGER;<br />

GATE SETTINGS: GATE MODE EDGE; POLARITY POS<br />

SWEEPTIME MANUAL {1} ms: GATE DELAY {300} µs:<br />

GATE LENGTH: {250} µs]<br />

The following figure shows the screen display for setting gate parameters. The vertical<br />

lines for gate delay (GD) and gate length (GL) can be adjusted to the burst signal<br />

by entering numeric values or by means of the rotary knob.<br />

4.38 Operating Manual 1313.9681.12 - 01