R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Example: "CALC:MARK:X:SLIM ON"<br />

'Switches the search limit function on for screen A.<br />

"CALC:MARK:X:SLIM:RIGH 20MHz"<br />

'Sets the right limit of the search range in screen A to 20 MHz.<br />

CALCulate Subsystem<br />

Characteristics: *RST value: - (is set to the right diagram border on switching on search limits)<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:X:SLIMits[:STATe] ON | OFF<br />

This command switches between a limited (ON) and unlimited (OFF) search range in the selected<br />

measurement window. The function is independent of the selection of a marker, i.e. the numeric suffix<br />

MARKer is irrelevant.<br />

If the time domain power measurement is active, this command limits the evaluation range on the<br />

trace.<br />

Example: "CALC:MARK:X:SLIM ON"<br />

'Switches on search limitation in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:X:SSIZe STANdard | POINts<br />

This command defines the marker step size when you change the position of a marker with the rotary<br />

knob.<br />

The numeric suffixes at CALCulate and MARKer are irrelevant.<br />

Parameter: STANdard<br />

The step size of the marker is one pixel.<br />

POINts<br />

The step size of the marker is one sweep point.<br />

The number of sweep points in a trace depends on the number of sweep points you<br />

have set.<br />

Example: "CALC:MARK:X:SSIZ POIN"<br />

'The step size of the marker is one sweep point.<br />

Characteristics: *RST value: STANdard<br />

SCPI: device-specific<br />

Mode: A<br />

The function is only available if the search limit for marker and delta marker is<br />

switched on (CALC:MARK:X:SLIM ON).<br />

6.52 Operating Manual 1313.9681.12 - 01